National Instruments Digital Camera NI 1450 Series User Manual

TM  
IMAQ  
NI 1450 Series Compact Vision  
System User Manual  
NI 1450 Series Compact Vision System User Manual  
June 2003 Edition  
Part Number 323610A-01  
 
 
Important Information  
Warranty  
The NI 1450 Series Compact Vision System is warranted against defects in materials and workmanship for a period of one year from the date  
of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves  
to be defective during the warranty period. This warranty includes parts and labor.  
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects  
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National  
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives  
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be  
uninterrupted or error free.  
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before  
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are  
covered by warranty.  
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical  
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent  
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.  
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.  
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF  
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF  
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR  
DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY  
THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including  
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments  
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover  
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or  
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,  
flood, accident, actions of third parties, or other events outside reasonable control.  
Copyright  
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,  
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National  
Instruments Corporation.  
Trademarks  
IMAQ, LabVIEW, National Instruments, NI, ni.com, NI Developer Zone, and NI-IMAQare trademarks of National Instruments  
Corporation.  
Product and company names mentioned herein are trademarks or trade names of their respective companies.  
Patents  
For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txtfile  
on your CD, or ni.com/patents.  
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS  
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF  
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN  
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT  
INJURY TO A HUMAN.  
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE  
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,  
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS  
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND  
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL  
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR  
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE  
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD  
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD  
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID  
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO  
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.  
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING  
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN  
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL  
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING  
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE  
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,  
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.  
 
Compliance  
FCC/Canada Radio Frequency Interference Compliance  
Determining FCC Class  
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC  
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)  
or Class B (for use in residential or commercial locations). All National Instruments (NI) products are FCC Class A products.  
Depending on where it is operated, this Class A product could be subject to restrictions in the FCC rules. (In Canada, the  
Department of Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.) Digital  
electronics emit weak signals during normal operation that can affect radio, television, or other wireless products.  
All Class A products display a simple warning statement of one paragraph in length regarding interference and undesired  
operation. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.  
Consult the FCC Web site at www.fcc.govfor more information.  
FCC/DOC Warnings  
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions  
in this manual and the CE marking Declaration of Conformity*, may cause interference to radio and television reception.  
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department of  
Communications (DOC).  
Changes or modifications not expressly approved by NI could void the user’s authority to operate the equipment under the FCC  
Rules.  
Class A  
Federal Communications Commission  
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC  
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated  
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and  
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this  
equipment in a residential area is likely to cause harmful interference in which case the user is required to correct the interference  
at their own expense.  
Canadian Department of Communications  
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.  
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.  
Compliance to EU Directives  
Readers in the European Union (EU) must refer to the manufacturer’s Declaration of Conformity (DoC) for information*  
pertaining to the CE marking compliance scheme. The manufacturer includes a DoC for most hardware products except for those  
bought from OEMs. In addition, DoCs are usually not provided if compliance is not required, for example electrically benign  
apparatus or cables.  
To obtain the DoC for this product, click Declarations of Conformity Information at ni.com/hardref.nsf/. This Web site  
lists the DoCs by product family. Select the appropriate product family, followed by your product, and a link to the DoC appears  
in Adobe Acrobat format. Click the Acrobat icon to download or read the DoC.  
*
The CE marking Declaration of Conformity contains important supplementary information and instructions for the user or  
installer.  
 
Conventions  
The following conventions are used in this manual:  
»
The » symbol leads you through nested menu items and dialog box options  
to a final action. The sequence File»Page Setup»Options directs you to  
pull down the File menu, select the Page Setup item, and select Options  
from the last dialog box.  
This icon denotes a tip, which alerts you to advisory information.  
This icon denotes a note, which alerts you to important information.  
This icon denotes a caution, which advises you of precautions to take to  
avoid injury, data loss, or a system crash.  
bold  
Bold text denotes items that you must select or click in the software, such  
as menu items and dialog box options. Bold text also denotes hardware  
labels and parameter names.  
italic  
Italic text denotes variables, emphasis, a cross reference, or an introduction  
to a key concept. This font also denotes text that is a placeholder for a word  
or value that you must supply.  
monospace  
Text in this font denotes text or characters that you should enter from the  
keyboard, sections of code, programming examples, and syntax examples.  
This font is also used for the proper names of disk drives, paths, directories,  
programs, subprograms, subroutines, device names, functions, operations,  
variables, filenames and extensions, and code excerpts.  
monospace bold  
Bold text in this font denotes the messages and responses that the computer  
automatically prints to the screen. This font also emphasizes lines of code  
that are different from the other examples.  
 
 
Chapter 1  
NI 1450 Overview  
Available Camera Bandwidth........................................................................................1-3  
Software Overview ........................................................................................................1-4  
National Instruments Application Software ....................................................1-5  
Chapter 2  
Software...........................................................................................................2-2  
Accessories ....................................................................................................................2-2  
Documentation...............................................................................................................2-3  
Hardware Documents ......................................................................................2-3  
Wiring Power to the NI 1450............................................................2-9  
Connecting the NI 1450 to the Development Computer...................2-11  
Vision Builder AI: Setting up the Development Computer...........................................2-12  
Installing Vision Builder AI and NI-IMAQ for IEEE 1394 Cameras.............2-13  
Configuring the IP Address and Downloading Software onto the NI 1450....2-13  
Acquiring an Image in Vision Builder AI.......................................................2-14  
© National Instruments Corporation  
vii  
NI 1450 Series Compact Vision System User Manual  
 
Contents  
Development Computer.............................................................................................. 2-15  
Installing LabVIEW Real-Time, Vision Development Module,  
Configuring the IP Address using LabVIEW Real-Time ............................... 2-16  
Chapter 3  
VGA................................................................................................................ 3-8  
COM1.............................................................................................................. 3-10  
TRIG 0 ............................................................................................................ 3-11  
Chapter 4  
Isolated Inputs and Outputs............................................................................. 4-2  
Trigger Inputs.................................................................................................. 4-3  
Timed Pulse Output ........................................................................................ 4-4  
Initiating a Timed Pulse.................................................................... 4-4  
Pulse Modes...................................................................................... 4-5  
Pulse Delay....................................................................................... 4-5  
NI 1450 Series Compact Vision System User Manual  
viii  
ni.com  
 
Considerations When Connecting the Digital I/O...........................................4-11  
Wiring an Isolated Input to a Sourcing Output Device.....................4-11  
Protecting Inductive Loads ...............................................................4-13  
Typical System Setup......................................................................................4-14  
Deployment  
Connecting Multiple NI 1450s ......................................................................................5-1  
Appendix A  
Troubleshooting  
Appendix B  
Specifications  
Appendix C  
Technical Support and Professional Services  
Glossary  
Index  
© National Instruments Corporation  
ix  
NI 1450 Series Compact Vision System User Manual  
 
1
NI 1450 Overview  
This chapter provides an overview of the features and components on the  
National Instruments 1450 Series Compact Vision System.  
About the NI 1450 Series Compact Vision System  
The NI 1450 Series Compact Vision System is an easy-to-use, distributed,  
real-time imaging system that acquires, processes, and displays images  
from IEEE 1394 cameras conforming to the IIDC 1394-based Digital  
Camera Specification, Version 1.30. The NI 1450 also provides multiple  
digital input/output (I/O) options for communicating with external devices  
to configure and start an inspection and to indicate results.  
An Ethernet connection between the NI 1450 and a development computer  
allows you to display measurement results and status information and to  
configure the NI 1450 settings. Once configured, the NI 1450 can run  
applications without a connection to the development computer.  
Each NI 1450 ships with documentation and the NI-IMAQ for IEEE 1394  
Cameras driver software, which is licensed for one development system  
and one deployment system.  
Hardware Overview  
The NI 1450 front panel consists of a VGA connector, an RS-232 serial  
port, a 10/100 Ethernet connector, and three IEEE 1394a ports.  
The NI 1450 also includes LEDs for communicating system status, DIP  
switches for mode control for specifying startup options, TTL inputs and  
outputs for triggering, and isolated inputs and outputs for connecting to  
external devices, such as PLCs, sensors, LED indicators, and start/stop  
buttons. The isolated inputs and outputs on the NI 1450 provide an easy  
means for preventing ground loops that could degrade signal integrity.  
Caution The isolation on the NI 1450 is not safety isolation.  
© National Instruments Corporation  
1-1  
NI 1450 Series Compact Vision System User Manual  
 
               
Chapter 1  
NI 1450 Overview  
Figure 1-1 shows the NI 1450 front panel.  
1
2
NI 1454  
Compact Vision System  
3
4
11  
10  
5
9
8
6
7
1
2
3
4
Power LED  
Status LED  
Isolated Digital Input  
TTL Digital Outputs  
5
6
7
8
IEEE 1394a Ports  
TTL I/O and Isolated I/O 10 RS-232 Serial  
Reset Button  
DIP Switches  
9
VGA  
11 RJ-45 Ethernet Port  
Figure 1-1. NI 1450 Series Compact Vision System Front Panel  
NI 1450 Series Compact Vision System User Manual  
1-2  
ni.com  
 
   
Chapter 1  
NI 1450 Overview  
Available Camera Bandwidth  
The IEEE 1394 bus provides a fixed amount of bandwidth that is shared  
among the three IEEE 1394a ports on the NI 1450. These ports provide  
direct connection to up to three DCAM-compliant IEEE 1394 cameras,  
depending on the amount of bandwidth each camera requires. Higher frame  
rates and larger image sizes require a higher data transfer rate and use more  
bandwidth.  
Table 1-1 shows the maximum number of cameras the NI 1450 supports for  
three common video formats. Use this table as a guide when determining  
the combination of cameras to use in your application. The maximum  
number of cameras listed in the table does not include processing time.  
By triggering your cameras slower than the frame rates listed in the table,  
you can operate more cameras simultaneously. If the camera combination  
exceeds the amount of available bandwidth, the software returns an  
Insufficient Resources error.  
Table 1-1. Available Camera Bandwidth  
Maximum Number  
of Cameras for  
Video Format  
Frames per Second Simultaneous Operation  
640 x 480,  
8-bit/pixel mono  
30  
60  
3
1
1
3
1
100  
15  
640 x 480 YUV  
(4:2:2) 16-bit/pixel  
color  
30  
1024 x 768  
16-bit/pixel mono  
7.5  
15  
2
1
© National Instruments Corporation  
1-3  
NI 1450 Series Compact Vision System User Manual  
 
           
Chapter 1  
NI 1450 Overview  
Software Overview  
Programming the NI 1450 Series Compact Vision System requires the  
NI-IMAQ for IEEE 1394 Cameras driver software, version 1.5 or later,  
to control the hardware and one of the following application software  
packages to process images.  
NI Vision Builder for Automated Inspection (AI), version 2.0 or  
laterTo configure solutions to common inspection tasks  
LabVIEW Real-Time (RT), version 7.0 or later, with the Vision  
Development Module, version 7.0 or later—For customizable  
control over acquisition hardware and algorithms  
The following sections provide an overview of the driver software and the  
application software. For detailed information about individual software  
packages, refer to the documentation specific to the package.  
NI-IMAQ for IEEE 1394 Cameras Driver Software  
The NI-IMAQ for IEEE 1394 Cameras driver software ships with the  
NI 1450 Series Compact Vision System. NI-IMAQ for IEEE 1394  
Cameras is the interface path between the application software and the  
NI 1450.  
NI-IMAQ for IEEE 1394 Cameras includes an extensive library of VIs  
you can call from LabVIEW. These VIs include routines for video  
control, and register-level camera configuration.  
The NI-IMAQ for IEEE 1394 Cameras driver software performs all  
functions necessary for acquiring and saving images but does not perform  
image analysis. For image analysis functionality, refer to the National  
Instruments Application Software section.  
For maximum flexibility and performance, NI-IMAQ for IEEE 1394  
Cameras features both high-level and low-level functions. A function that  
acquires images in multi-buffer, single-shot, or continuous mode is an  
example of a high-level function. A function that requires advanced  
understanding of the NI 1450 and image acquisition, such as configuring  
an image sequence, is an example of a low-level function.  
The NI-IMAQ for IEEE 1394 Cameras software handles many of the  
complex issues between the NI 1450 and the camera, such as 1394 bus  
communication and camera control.  
NI 1450 Series Compact Vision System User Manual  
1-4  
ni.com  
 
         
Chapter 1  
NI 1450 Overview  
National Instruments Application Software  
Vision Builder for Automated Inspection  
National Instruments Vision Builder for Automated Inspection (AI) is  
configurable machine vision software that you can use to prototype,  
benchmark, and deploy applications. NI Vision Builder AI does not require  
programming, but is scalable to LabVIEW RT.  
Vision Builder AI allows you to easily configure and benchmark a  
sequence of visual inspection steps, as well as deploy the visual inspection  
system for automated inspection. With Vision Builder AI, you can perform  
powerful visual inspection tasks and make decisions based on the results of  
individual tasks. Additionally, you can communicate with external devices  
using an Ethernet connection, serial lines, and digital I/Os. With Vision  
Builder AI, you can migrate your configured inspection to LabVIEW,  
extending the capabilities of your applications if necessary.  
LabVIEW RT with the Vision Development Module  
The LabVIEW Real-Time Module and the RT Series hardware, such as the  
NI 1450, extend the capabilities of LabVIEW to address the need for  
deterministic real-time performance.  
The Real-Time Module combines LabVIEW graphical programming with  
the power of RT Series hardware, enabling you to build deterministic  
real-time systems. You develop VIs in LabVIEW and embed the VIs on  
RT targets. The RT target runs VIs without a user interface and offers a  
stable platform for real-time VIs.  
For more information about LabVIEW RT, refer to the LabVIEW  
Real-Time Module User Manual.  
The Vision Development Module is an image acquisition, processing, and  
analysis library of more than 270 functions for grayscale, color, and binary  
image display, image processing, pattern matching, particle analysis,  
gauging, and measurement.  
For unique image processing, you can use the Vision Development Module  
functions individually or in combination. The Vision Development Module  
allows you to acquire, display, manipulate, and store images as well as  
perform image analysis, processing, and interpretation. Using the Vision  
Development Module, imaging novices and experts can program the most  
basic or complicated image applications without knowledge of particular  
algorithm implementations.  
© National Instruments Corporation  
1-5  
NI 1450 Series Compact Vision System User Manual  
 
             
Chapter 1  
NI 1450 Overview  
Vision Assistant is an interactive prototyping tool for machine vision and  
scientific imaging developers and is included with the Vision Development  
Module. With Vision Assistant, you can prototype vision applications  
quickly and test how various vision image processing functions work.  
For information about how to use the Vision Development Module with  
LabVIEW RT, refer to the IMAQ Vision for LabVIEW User Manual.  
NI 1450 Series Compact Vision System User Manual  
1-6  
ni.com  
 
2
Setup and Configuration  
This chapter provides instructions for connecting the NI 1450 Series  
Compact Vision System hardware. This chapter also includes instructions  
for installing software, configuring an IP address, and acquiring an initial  
image using the application software.  
Required Items  
The following items are necessary for connecting the NI 1450.  
Hardware  
NI 1450 Series Compact Vision System  
Ethernet-equipped development computer running  
Windows XP/2000/Me/98  
DCAM-compliant IEEE 1394 camera  
IEEE 1394 jackscrew-to-latch cable (part number 778796-01) or any  
standard IEEE 1394 cable—for plug-and-play connection from the  
NI 1450 to up to three 1394 cameras. You can use a 4-pin to 6-pin  
converter cable with cameras that have their own external power  
supply and do not require power from the 1394 bus.  
Note To maintain signal integrity, the IEEE 1394 cable length must be no longer  
than 4.5 m.  
NI desktop power supply (part number 778794-01) or any  
24 VDC 10%, 50 W power supply  
Power supply cord—for connecting the NI desktop power supply to an  
outlet. Refer to ni.comfor ordering information for the part number  
specific to your region.  
Note If you are using the NI desktop power supply, you will need a power cord.  
© National Instruments Corporation  
2-1  
NI 1450 Series Compact Vision System User Manual  
 
                 
Chapter 2  
Setup and Configuration  
4-position power connector—required if you are not using the  
NI desktop power supply  
10 m 10/100Base-T Ethernet cable (part number  
189174-10)—standard CAT 5 10/100Base-T Ethernet cable for  
connecting the NI 1450 to a network port. To connect the NI 1450  
directly to a local development computer, use an Ethernet crossover  
cable.  
Note To maintain signal integrity, the Ethernet cable length must be no longer than 100 m.  
Software  
NI-IMAQ for IEEE 1394 Cameras driver software, which includes the  
NI 1450 support software  
One of the following application software packages:  
National Instruments Vision Builder for Automated Inspection  
(AI), version 2.0 or later  
National Instruments LabVIEW Real-Time, version 7.0 or later,  
with the Vision Development Module, version 7.0 or later  
Accessories  
National Instruments offers the following accessories for use with the  
NI 1450.  
Hardware  
VGA-resolution camera (part number 778785-01)  
12 mm fixed focal length lens (part number 778789-01)  
LED ring light (part number 778787-01)  
DIN rail/panel mount kit (part number 189154-01)  
Digital I/O cable and horizontal DIN rail terminal block (part number  
778790-01)  
NI 1450 Series Compact Vision System User Manual  
2-2  
ni.com  
 
                 
Chapter 2  
Setup and Configuration  
Digital I/O cable and vertical DIN rail terminal block (part number  
778791-01)  
75 SMB 111 coaxial cable (part number 763422-01)— SMB to  
BNC cable for connecting to triggers and light sources  
10 m Ethernet crossover cable (part number 187375-10)—for  
connecting the NI 1450 directly to the development computer  
Documentation  
The documentation resources listed in this section ship with the NI 1450  
and the application software. For more resources, including code examples  
and tutorials, visit the National Instruments Developer Zone at  
ni.com/zone.  
Hardware Documents  
NI 1450 Series Compact Vision System User Manual  
NI 1450 Series Compact Vision System Digital I/O Help  
Vision Builder AI Documents  
NI Vision Builder for Automated Inspection Tutorial  
NI Vision Builder for Automated Inspection: Configuration Help  
NI Vision Builder for Automated Inspection: Inspection Help  
LabVIEW Real-Time Module with the Vision Development Module  
Documents  
NI-IMAQ for IEEE 1394 Cameras Driver Software  
Documents  
Getting Started with NI-IMAQ for IEEE 1394 Cameras  
NI-IMAQ for IEEE 1394 Cameras User Manual (PDF)  
NI-IMAQ for IEEE 1394 Cameras VI Reference Help  
MAX Help for NI-IMAQ for IEEE 1394 Cameras  
© National Instruments Corporation  
2-3  
NI 1450 Series Compact Vision System User Manual  
 
                   
Chapter 2  
Setup and Configuration  
LabVIEW Real-Time Module and Vision  
Development Module Documents  
NI Vision Assistant Help  
NI Vision Assistant Tutorial  
IMAQ Vision Concepts Manual  
IMAQ Vision for LabVIEW User Manual  
IMAQ Vision for LabVIEW Help  
LabVIEW Real-Time Module documentation  
Safety Information  
Caution The following paragraphs contain important safety information you must follow  
when installing and operating the device.  
Do not operate the device in a manner not specified in the documentation.  
Misuse of the device may result in a hazard and may compromise the safety  
protection built into the device. If the device is damaged, turn it off and do  
not use it until service-trained personnel can check its safety. If necessary,  
return the device to National Instruments for repair.  
Keep away from live circuits. Do not remove equipment covers or shields  
unless you are trained to do so. If signal wires are connected to the device,  
hazardous voltages can exist even when the equipment is turned off. To  
avoid a shock hazard, do not perform procedures involving cover or shield  
removal unless you are qualified to do so. Disconnect all field power prior  
to removing covers or shields.  
If the device is rated for use with hazardous voltages (>30 Vrms, 42.4 Vpk,  
or 60 Vdc), it may require a safety earth-ground connection wire. Refer to  
the device specifications for maximum voltage ratings.  
Because of the danger of introducing additional hazards, do not install  
unauthorized parts or modify the device. Use the device only with the  
chassis, modules, accessories, and cables specified in the installation  
instructions. All covers and filler panels must be installed while operating  
the device.  
Do not operate the device in an explosive atmosphere or where flammable  
gases or fumes may be present. Operate the device only at or below the  
pollution degree stated in the specifications. Pollution consists of any  
NI 1450 Series Compact Vision System User Manual  
2-4  
ni.com  
 
     
Chapter 2  
Setup and Configuration  
foreign matter—solid, liquid, or gas—that may reduce dielectric strength  
or surface resistivity. Pollution degrees are listed below.  
Pollution Degree 1—No pollution or only dry, nonconductive  
pollution occurs. The pollution has no effect.  
Pollution Degree 2—Normally only nonconductive pollution occurs.  
Occasionally, nonconductive pollution becomes conductive because of  
condensation.  
Pollution Degree 3—Conductive pollution or dry, nonconductive  
pollution occurs. Nonconductive pollution becomes conductive  
because of condensation.  
Clean the device and accessories by brushing off light dust with a soft,  
nonmetallic brush. Remove other contaminants with a stiff, nonmetallic  
brush. The unit must be completely dry and free from contaminants before  
returning it to service.  
You must insulate signal connections for the maximum voltage for which  
the device is rated. Do not exceed the maximum ratings for the device.  
Remove power from signal lines before connection to or disconnection  
from the device.  
Caution National Instruments measurement products may be classified as either  
Installation Category I or II. Operate products at or below the Installation Category level  
specified in the hardware specifications.  
Installation Category1: Measurement circuits are subjected to working  
voltages2 and transient stresses (overvoltage) from the circuit to which they  
are connected during measurement or test. Installation Category establishes  
standardized impulse withstand voltage levels that commonly occur in  
electrical distribution systems. The following is a description of Installation  
(Measurement3) Categories:  
Installation Category I is for measurements performed on circuits not  
directly connected to the electrical distribution system referred to as  
MAINS4 voltage. This category is for measurements of voltages from  
specially protected secondary circuits. Such voltage measurements  
include signal levels, special equipment, limited-energy parts of  
1
2
3
4
Installation Categories as defined in electrical safety standard IEC 61010-1.  
Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.  
Installation Category is also referred to as Measurement Category.  
MAINS is defined as the (hazardous live) electrical supply system to which equipment is deisgned to be connected for the  
purpose of powering the equipment. Suitably rated measuring circuits may be connected to the MAINS for measuring  
purposes.  
© National Instruments Corporation  
2-5  
NI 1450 Series Compact Vision System User Manual  
 
 
Chapter 2  
Setup and Configuration  
equipment, circuits powered by regulated low-voltage sources, and  
electronics.  
Installation Category II is for measurements performed on circuits  
directly connected to the electrical distribution system. This category  
refers to local-level electrical distribution, such as that provided by a  
standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe).  
Examples of Installation Category II are measurements performed on  
household appliances, portable tools, and similar products.  
Installation Category III is for measurements performed in the building  
installation at the distribution level. This category refers to  
measurements on hard-wired equipment such as equipment in fixed  
installations, distribution boards, and circuit breakers. Other examples  
are wiring, including cables, bus-bars, junction boxes, switches,  
socket-outlets in the fixed installation, and stationary motors with  
permanent connections to fixed installations.  
Installation Category IV is for measurements performed at the primary  
electrical supply installation (<1,000 V). Examples include electricity  
meters and measurements on primary overcurrent protection devices  
and on ripple control units.  
Connection Overview  
Figure 2-1 illustrates the sequence for connecting and getting started with  
the NI 1450.  
NI 1450 Series Compact Vision System User Manual  
2-6  
ni.com  
 
   
Chapter 2  
Setup and Configuration  
Set Up the Hardware  
Set Up the Development  
Computer Using LabVIEW  
Real-Time with the Vision  
Development Module  
Set Up the Development  
Computer Using  
Vision Builder AI  
or  
or  
Acquire an Image Using  
LabVIEW Real-Time with the  
Vision Development Module  
Acquire an Image Using  
Vision Builder AI  
Figure 2-1. Connection Sequence  
Set up the hardware—This section explains how to connect a camera,  
monitor, and power supply to the NI 1450.  
Set up the development computer—This section explains how to use  
either Vision Builder AI or LabVIEW Real-Time with the Vision  
Development Module to perform the following tasks:  
Connect the NI 1450 to the development computer  
Install application and driver software  
Obtain an IP address  
Acquire an image—This section explains how to use either Vision  
Builder AI or LabVIEW Real-Time with the Vision Development  
Module to acquire an image.  
Before Getting Started: Connecting the NI 1450 to a Network  
Use a standard Category 5 or Category 6 Ethernet cable to connect the  
NI 1450 to an Ethernet network.  
If the development computer is already configured on a network, you must  
configure the NI 1450 on the same network. If the development computer  
is not connected to a network, you can connect the two directly using a  
Category 5 or Category 6 crossover cable.  
Caution To prevent data loss and to maintain the integrity of your Ethernet installation,  
do not use a cable longer than 100 m. If you are using a 100 Mbps Ethernet, National  
Instruments recommends using a Category 5 or Category 6 shielded twisted-pair  
Ethernet cable.  
© National Instruments Corporation  
2-7  
NI 1450 Series Compact Vision System User Manual  
 
     
Chapter 2  
Setup and Configuration  
Subnet Considerations  
To configure the NI 1450, it must reside on the same subnet as the  
development computer. Once the NI 1450 is configured, other subnets can  
access and use it.  
To use the NI 1450 on a subnet other than the one the development  
computer is on, first connect and configure it on the same subnet as the  
development computer. Next, physically move it to the other subnet and  
reassign an IP address. Contact your network administrator for assistance  
in setting up the development computer and NI 1450 on the same subnet.  
Hardware Setup  
This section describes how to connect the basic hardware components of  
the NI 1450. When these basic components are connected for the first time,  
the NI 1450 runs a program that acquires images. This program verifies that  
all hardware components are properly connected and functioning.  
The following items are necessary for hardware setup.  
NI 1450 Series Compact Vision System  
24 VDC 10%, 50 W power supply  
DCAM-compatible IEEE 1394 camera  
IEEE 1394 cable  
Ethernet cable  
Monitor  
Connecting a Camera and Monitor to the NI 1450  
Before connecting a camera and monitor to the NI 1450, make sure that all  
NI 1450 DIP switches are in the OFF position.  
To connect an IEEE 1394 camera and a monitor to the NI 1450, refer to  
Figure 2-2 while completing the following steps:  
1. Connect the VGA cable from the monitor to the VGA port on the  
NI 1450.  
2. Plug the IEEE 1394 cable into one of the IEEE 1394a ports on the  
NI 1450. Plug the other end of the cable into the IEEE 1394 receptacle  
on the camera.  
NI 1450 Series Compact Vision System User Manual  
2-8  
ni.com  
 
               
Chapter 2  
Setup and Configuration  
If your camera requires an external power supply, connect it to the  
camera and verify that the camera is powered on.  
3. Plug in and power on the monitor.  
1
2
1
VGA Cable  
2
IEEE 1394 Cable  
Wiring Power to the NI 1450  
This section describes how to connect the NI desktop power supply.  
For instructions on how to connect a separate main supply, refer to the  
Connecting to a Separate Main Supply section.  
Caution Do not connect the NI 1450 main power to a source other than 24 VDC 10%.  
Do not connect the NI 1450 isolated power to a source less than 5 VDC or greater than  
30 VDC. Doing so could damage the NI 1450.  
To connect power to the NI 1450, refer to Figure 2-3 while completing the  
following steps:  
1. Plug the 4-position connector from the power supply into the power  
receptacle on the NI 1450.  
2. Plug the power cord into the power supply.  
3. Plug the power cord into an outlet.  
© National Instruments Corporation  
2-9  
NI 1450 Series Compact Vision System User Manual  
 
     
Chapter 2  
Setup and Configuration  
The NI 1450 ships with a factory-installed startup program that, when the  
NI 1450 is connected to a camera and powered on, acquires images and  
displays them on the monitor. If these images from the camera display on  
the monitor, continue to the Connecting the NI 1450 to the Development  
Computer section. If the images from the camera are not displayed on the  
monitor, refer to Appendix A, Troubleshooting.  
To Outlet  
1
2
3
1
4-Position Power Connector  
2
NI Desktop Power Supply  
3
Power Supply Cord (to Outlet)  
Figure 2-3. Wiring Power to the NI 1450  
Connecting to a Separate Main Supply  
If you are using a power supply other than the NI desktop power supply,  
use this section to connect power to the NI 1450.  
Caution Do not connect the NI 1450 main power to a source other than 24 VDC 10%.  
Do not connect the NI 1450 isolated power to a source less than 5 VDC or greater than  
30 VDC. Doing so could damage the NI 1450.  
NI 1450 Series Compact Vision System User Manual  
2-10  
ni.com  
 
       
Chapter 2  
Setup and Configuration  
The NI 1450 ships with a 4-position power connector that plugs directly  
into the power input connector on the NI 1450. To wire power to the  
4-position connector, complete the following steps:  
1. Wire the voltage output of the 24 VDC 10% power supply to the main  
voltage input, labeled V, on the 4-position connector.  
2. Wire the common (ground) output of the power supply to the common  
input, labeled C, on the 4-position connector.  
If you are using a separate power supply for the NI 1450 isolated outputs,  
connect the voltage output on the power supply to the isolated power (Viso)  
on the 4-position connector. Connect the common (ground) on the power  
supply to the isolated common (Ciso) on the connector.  
Note If you do not require a separate power supply for the NI 1450 isolated outputs,  
you can daisy-chain the V to the Viso and the C to the Ciso on the connector.  
For information about grounding the NI 1450 chassis to earth ground, refer  
to the Earth Ground Connection section of Chapter 3, LEDs, DIP Switches,  
and Connectors.  
Connecting the NI 1450 to the Development  
Computer  
The development computer communicates with the NI 1450 over an  
Ethernet connection. Use a standard Ethernet cable to connect from the  
network port to the NI 1450. To connect the NI 1450 directly to the  
development computer, use an Ethernet crossover cable.  
To connect the NI 1450 to the development computer, refer to Figure 2-4  
while completing the following steps:  
1. Verify that the development computer is connected to the network and  
is powered on.  
2. Using a standard CAT 5 Ethernet cable, connect from the network port  
to the Ethernet port on the NI 1450.  
3. Using a standard CAT 5 Ethernet cable, connect from the network port  
to the Ethernet port on the development computer.  
Note If you are not connecting through a network, use an Ethernet crossover cable to  
connect the NI 1450 directly to the development computer.  
© National Instruments Corporation  
2-11  
NI 1450 Series Compact Vision System User Manual  
 
   
Chapter 2  
Setup and Configuration  
1
2
1
2
3
Standard Ethernet Cable Connecting from the NI 1450 to an Ethernet Hub  
Standard Ethernet Cable Connecting from an Ethernet Hub to the Development Computer  
Ethernet Hub or Other Network Port  
Figure 2-4. Ethernet Connection  
Vision Builder AI: Setting up the Development Computer  
This section describes the sequence for installing Vision Builder AI and the  
obtaining an IP address, installing software on the NI 1450, and  
configuring the NI 1450 to acquire an image using Vision Builder AI.  
This section applies only to Vision Builder AI users. If you are using  
LabVIEW RT with the Vision Development Module, refer to the LabVIEW  
Real-Time with the Vision Development Module: Setting up the  
Development Computer section.  
NI 1450 Series Compact Vision System User Manual  
2-12  
ni.com  
 
       
Chapter 2  
Setup and Configuration  
The following items are necessary for setting up the development  
computer.  
Vision Builder AI software, version 2.0 or later  
NI-IMAQ for IEEE 1394 Cameras driver software, version 1.5 or later  
Installing Vision Builder AI and NI-IMAQ for IEEE 1394 Cameras  
This section describes how to install the Vision Builder AI software and the  
NI-IMAQ for IEEE 1394 Cameras driver software onto the development  
computer.  
Note You must install Vision Builder AI before installing the NI-IMAQ for IEEE 1394  
Cameras driver software.  
Complete the following steps to install Vision Builder AI and the NI-IMAQ  
for IEEE 1394 Cameras driver software onto the development computer.  
1. Insert the Vision Builder AI CD into the CD-ROM drive.  
2. When the installation splash screen appears, click Install NI Vision  
Builder AI 2.0 and follow the setup instructions.  
3. Insert the NI-IMAQ for IEEE 1394 Cameras CD into the CD-ROM  
drive.  
Note When installing NI-IMAQ for IEEE 1394 Cameras, make sure to install the support  
files for LabVIEW Real-Time.  
4. When the installation splash screen appears, click Install NI-IMAQ  
for IEEE 1394 Cameras and follow the setup instructions.  
5. Reboot the development computer.  
Configuring the IP Address and Downloading Software onto the NI 1450  
To set up an IP address and download software onto the NI 1450, complete  
the following steps:  
1. Launch Vision Builder AI by navigating to Start»Programs»  
National Instruments»Vision Builder AI.  
2. Expand the Execution Target pull-down listbox and click Select  
Network Target. This process takes several seconds.  
3. In the Select Remote Target window, click 192.168.10.12 to highlight  
the row. This IP address is assigned to all unconfigured NI 1450s.  
© National Instruments Corporation  
2-13  
NI 1450 Series Compact Vision System User Manual  
 
         
Chapter 2  
Setup and Configuration  
Tip To uniquely identify unconfigured NI 1450s, connect and configure one NI 1450  
at a time.  
4. Click Configure to launch the Vision Builder AI Remote Target  
Configuration Wizard.  
5. In the Identification window, enter a name for the NI 1450 in the  
Name field and a description of the NI 1450 in the Description field.  
Note Device names are limited to 15 characters with no spaces or special characters.  
The first and last characters must be alphanumeric.  
6. Click Next.  
7. If the network is configured to issue IP addresses using DHCP, select  
Obtain IP address from DHCP server. Otherwise, set the IP address  
manually by selecting Edit IP Settings, Suggest Values, and OK.  
8. Click Next. This window shows the status of the software installed on  
the NI 1450.  
9. Select the Update Target Software checkbox.  
10. Click OK to begin configuring the IP address and downloading  
software onto the NI 1450. This initialization process takes several  
minutes.  
Acquiring an Image in Vision Builder AI  
To acquire an image in Vision Builder AI, complete the following steps:  
1. Launch Vision Builder AI.  
2. In the Vision Builder AI Welcome screen, expand the Execution  
Target pull-down listbox and click Select Network Target.  
3. Select the NI 1450 you configured and click OK.  
4. In the Vision Builder AI Welcome screen, click Configure  
Inspection.  
5. From the Acquire Images palette, click Acquire Image (IEEE 1394).  
6. Click the Snap button to acquire a single image, or click the Grab  
button to acquire continuous images.  
7. Once you have configured your acquisition, click OK to add the step.  
You can now add inspection steps as documented in the NI Vision  
Builder for Automated Inspection Tutorial.  
NI 1450 Series Compact Vision System User Manual  
2-14  
ni.com  
 
   
Chapter 2  
Setup and Configuration  
LabVIEW Real-Time with the Vision Development  
Module: Setting up the Development Computer  
This section describes the sequence for installing the application and driver  
software on the development computer, obtaining an IP address, installing  
software on the NI 1450, and configuring the NI 1450 to acquire an image  
using LabVIEW Real-Time.  
This section applies only to LabVIEW Real-Time users. If you are using the  
Vision Builder AI software, refer to the Vision Builder AI: Setting up the  
Development Computer section.  
The following items are necessary for setting up the development  
computer.  
LabVIEW and LabVIEW Real-Time software  
Vision Development Module software  
NI-IMAQ for IEEE 1394 Cameras driver software  
Installing LabVIEW Real-Time, Vision Development Module,  
and NI-IMAQ for IEEE 1394 Cameras  
Note You must install LabVIEW, LabVIEW Real-Time, and the Vision Development  
Module software before installing the NI-IMAQ for IEEE 1394 Cameras driver software.  
Complete the following steps to install LabVIEW, LabVIEW Real-Time,  
the Vision Development Module, and the NI-IMAQ for IEEE 1394  
Cameras software onto the development computer.  
1. Insert the LabVIEW CD into the CD-ROM drive.  
2. When the installation splash screen appears, click Install LabVIEW  
and follow the setup instructions.  
3. Insert the LabVIEW Real-Time CD into the CD-ROM drive.  
4. When the installation splash screen appears, click Install LabVIEW  
Real-Time and follow the setup instructions.  
5. Insert the Vision Development Module CD into the CD-ROM drive.  
6. When the splash screen appears, click Install Vision 7.0 Development  
Module and follow the setup instructions.  
© National Instruments Corporation  
2-15  
NI 1450 Series Compact Vision System User Manual  
 
       
Chapter 2  
Setup and Configuration  
7. Insert the NI-IMAQ for IEEE 1394 Cameras CD into the CD-ROM  
drive.  
Note If you select the custom software installation, make sure to install the support for the  
NI 1450 Series Compact Vision System.  
8. When the splash screen appears, click Install NI-IMAQ for  
IEEE 1394 Cameras and follow the setup instructions.  
9. When prompted, click Yes to reboot the development computer.  
Configuring the IP Address using LabVIEW Real-Time  
To set up an IP address for the NI 1450, complete the following steps:  
1. Open the Measurement & Automation Explorer (MAX) configuration  
software by double-clicking the MAX icon on the desktop, or navigate  
to it by selecting Start»Programs»National Instruments»  
Measurement & Automation.  
2. Expand the Remote Systems branch of the configuration tree, and  
click 192.168.10.12 to display the Network Settings window.  
This IP address is assigned to all unconfigured NI 1450s.  
Tip To uniquely identify unconfigured NI 1450s, connect and configure one NI 1450  
at a time.  
3. In the Network Settings window, enter a name for the device in the  
Name field and a description of the device in the Comment field.  
Note Device names are limited to 15 characters with no spaces or special characters.  
The first and last characters must be alphanumeric.  
4. If the network is configured to issue IP addresses using DHCP, select  
Obtain IP address from DHCP server. Otherwise, set the IP address  
manually by selecting Edit the IP settings, Suggest Values, and OK.  
5. Click Apply.  
6. When prompted, click Yes to reboot the NI 1450. This initialization  
process takes several minutes.  
While the NI 1450 is rebooting, an icon  
appears next to the device  
name to indicate that the NI 1450 is disconnected. The MAX status bar  
also indicates the connection status of the NI 1450.  
NI 1450 Series Compact Vision System User Manual  
2-16  
ni.com  
 
     
Chapter 2  
Setup and Configuration  
Downloading Software onto the NI 1450  
1. In the MAX configuration window, click the Software tab. This  
window displays the status of the software on the NI 1450.  
2. Click the Install Software button.  
3. Select the software to download. For initial installation, make sure all  
checkboxes are selected.  
4. Click OK.  
5. When prompted, click Yes to reboot the NI 1450. This process takes  
several seconds.  
Acquiring an Image Using LabVIEW Real-Time  
To acquire an image using LabVIEW Real-Time, complete the following  
steps:  
1. Launch LabVIEW by navigating to Start»Programs»National  
Instruments LabVIEW 7.0.  
2. Expand the Execution Target pull-down listbox and click Select  
Target with Options.  
3. Enter the new IP address in the Machine Name/IP field and click OK.  
4. Click the Open button.  
5. Navigate to ProgramFiles\NationalInstruments\  
LabVIEW7.0\examples\IMAQ.  
6. Double-click IMAQ1394examples.llband select Grab.vi.  
7. Click the Run button to begin acquiring images.  
Now that you are acquiring images in LabVIEW, you can use the Vision  
Development Module and the installed NI 1450 drivers to process images  
and to control inputs and outputs.  
© National Instruments Corporation  
2-17  
NI 1450 Series Compact Vision System User Manual  
 
     
3
LEDs, DIP Switches, and  
This chapter provides information about the location and functionality of  
the LED indicators, DIP switches, and connectors on the NI 1450. The  
Connectors section provides signal names and descriptions for each  
connector.  
LED Indicators  
Figure 3-1 shows the location of the POWER OK and STATUS LEDs on  
the NI 1450.  
STATUS  
NI 1454  
Compact Vision System  
POWER OK  
Figure 3-1. POWER OK and STATUS LEDs  
Refer to Appendix A, Troubleshooting, for information about  
troubleshooting LEDs.  
© National Instruments Corporation  
3-1  
NI 1450 Series Compact Vision System User Manual  
 
           
Chapter 3  
LEDs, DIP Switches, and Connectors  
POWER OK LED  
Under normal operating conditions, the POWER OK LED remains green  
while the NI 1450 is powered on. A green POWER OK LED indicates that  
NI 1450 main power is receiving power and that the NI 1450 is not in a fault  
state. A red POWER OK LED indicates that the NI 1450 has shut down  
because of a fault state. A fault state occurs when the user shutdown input  
is asserted, the processor overheats, or the watchdog timer expires.  
Note The POWER OK LED does not indicate the status of the isolated power, Viso.  
STATUS LED  
The orange STATUS LED remains off under normal operating conditions  
and flashes a specific number of times to indicate error conditions or certain  
DIP switch settings. The STATUS LED remains lit if the NI 1450 detects  
an internal error.  
Refer to the Hardware Errors section of Appendix A, Troubleshooting,  
for information about LED error indications.  
ACT/LINK LED  
100 Mbps LED  
The orange ACT/LINK LED blinks when the NI 1450 receives data from  
or transmits data to the network through the Ethernet connection. Unrelated  
network activity causes this LED to blink occasionally even when the  
NI 1450 is inactive.  
Figure 3-2 shows the location of the ACT/LINK LED on the NI 1450.  
The green 100 Mbps LED is lit when the network provides 100 Mbps  
support and the NI 1450 is communicating at 100 Mbps. If the 100 Mbps  
LED is not lit, the NI 1450 is not operating at 100 Mbps.  
Figure 3-2 shows the location of the 100 Mbps LED on the NI 1450.  
NI 1450 Series Compact Vision System User Manual  
3-2  
ni.com  
 
               
Chapter 3  
LEDs, DIP Switches, and Connectors  
NI 1454  
Compact Vision System  
Figure 3-2. ACT/LINK and 100 Mbps LEDs  
DIP Switches  
This section describes the SAFE MODE, IP RESET, NO APP, and  
USER 1 DIP switches on the NI 1450.  
To enable a DIP switch, move it to the ON (left) position and then reset the  
NI 1450 by pressing the RESET button for at least two seconds.  
Note You must reset the NI 1450 in order for the setting change to occur.  
© National Instruments Corporation  
3-3  
NI 1450 Series Compact Vision System User Manual  
 
       
Chapter 3  
LEDs, DIP Switches, and Connectors  
Figure 3-3 shows the location of the DIP switches on the NI 1450.  
NI 1454  
Compact Vision System  
ON  
Figure 3-3. DIP Switches  
SAFE MODE Switch  
To boot the NI 1450 in Safe mode, move the SAFE MODE switch to the  
ON position and reset the NI 1450. Use safe mode to reconfigure TCP/IP  
settings and to download or update software from the development  
computer.  
Downloading incorrect software to the NI 1450 may cause it to hang on  
reboot or become inaccessible over the network. Powering on or resetting  
the NI 1450 in SAFE MODE starts the NI 1450 but does not start the  
embedded LabVIEW RT engine. To resume normal operations, reboot the  
NI 1450 with the SAFE MODE switch in the OFF position.  
NI 1450 Series Compact Vision System User Manual  
3-4  
ni.com  
 
       
Chapter 3  
LEDs, DIP Switches, and Connectors  
IP RESET Switch  
To clear the NI 1450 IP settings, move the IP RESET switch to the ON  
position and reset the NI 1450. Use IP RESET to reset the TCP/IP settings  
when moving the system from one subnet to another or when the current  
TCP/IP settings are invalid.  
Resetting the NI 1450 with the IP RESET switch in the ON position resets  
the IP address to 0.0.0.0. You can then set up a new network configuration  
for the NI 1450 from a development machine on the same subnet, or you  
can use an Ethernet crossover cable to connect the NI 1450 directly to the  
development computer.  
NO APP Switch  
To prevent the NI 1450 from automatically running VIs at startup, move  
the NO APP switch to the ON position and reset the NI 1450. If the  
NI 1450 becomes inaccessible because of a startup program, enable the  
NO APP switch and reset the NI 1450.  
Enable this switch to prevent the NI 1450 default startup program or Vision  
Builder AI from running at startup.  
USER 1 Switch (LabVIEW RT Users)  
The USER 1 switch is user-configurable and has no default functionality.  
You can use the RT Read Switch VI to read the USER 1 switch state and  
perform a custom action based on the current switch state position.  
© National Instruments Corporation  
3-5  
NI 1450 Series Compact Vision System User Manual  
 
         
Chapter 3  
LEDs, DIP Switches, and Connectors  
Connectors  
This section describes the connectors on the NI 1450 and includes pinouts  
and signal descriptions for each connector.  
Table 3-1 summarizes the functions of the connectors on the NI 1450.  
Table 3-1. NI 1450 Connectors Overview  
Peripheral  
External Connectors  
Function  
Power  
4-position power  
connector  
Main power and power for isolated outputs  
IEEE 1394a  
VGA  
6-pin IEEE 1394  
Power and data connection to IEEE 1394  
cameras  
15-pin female DSUB  
(standard VGA)  
Video output  
Serial  
9-pin male DSUB  
(standard RS-232 serial  
port)  
COM1  
10/100 Ethernet  
RJ-45 (standard  
Ethernet port)  
Ethernet network connection  
TRIG 0  
SMB receptacle  
SMB receptacle  
External isolated trigger input  
External TTL output  
TRIG 1 and TRIG 2  
Digital Input/Output  
44-pin female  
External TTL I/O; External isolated I/O  
high-density DSUB  
Power Connector  
The power connector on the NI 1450 accommodates two power supplies.  
The terminals labeled V and C provide the voltage and common for the  
main power of the NI 1450. The terminals labeled Viso and Ciso provide  
the voltage and common to power the isolated output circuitry.  
Caution The isolation provided by the NI 1450 is intended to prevent ground loops that  
could introduce noise into the system. This isolation does not provide safety isolation.  
NI 1450 Series Compact Vision System User Manual  
3-6  
ni.com  
 
                               
Chapter 3  
LEDs, DIP Switches, and Connectors  
Figure 3-4 shows the power connector on the NI 1450, and Table 3-2  
describes each terminal on the connector.  
POWER  
POWER  
Ciso  
Ciso  
Viso  
Viso  
(5-30 VDC)  
(5-30 VDC)  
C
V
(24 VDC 10%)  
C
V
(24 VDC 10%)  
Figure 3-4. Power Connector  
Table 3-2. Power Connector Terminals  
Terminal  
Description  
Main power (24 VDC 10%)  
Common  
V
C
Viso  
Ciso  
Isolated power (5 to 30 VDC)  
Isolated common  
Earth Ground Connection  
Some system setups may require using the grounding lug on the NI 1450 to  
connect the chassis to earth ground. Connecting the grounding lug, shown  
in Figure 3-5, to earth ground connects the common of the main power to  
earth ground through the NI 1450 chassis.  
Note An earth ground connection does not connect Ciso to earth ground.  
© National Instruments Corporation  
3-7  
NI 1450 Series Compact Vision System User Manual  
 
             
Chapter 3  
LEDs, DIP Switches, and Connectors  
1
POWER  
Ciso  
Viso  
(5-30VDC)  
2
C
V
(24VDC 10%)  
1
Grounding Lug  
2
Power Connector  
Figure 3-5. Grounding Lug on the NI 1450  
IEEE 1394  
The IEEE 1394 connectors on the NI 1450 provide a reliable,  
high-frequency connection between the NI 1450 and to up to  
three DCAM-compatible IEEE 1394 cameras. For information about  
the amount of bandwidth available for connecting cameras, refer to the  
Available Camera Bandwidth section of Chapter 1, NI 1450 Overview.  
To access the IEEE 1394 connectors on the NI 1450, use any standard 6-pin  
IEEE 1394 cable.  
Note You can use a 4-pin to 6-pin converter cable with cameras that have their own  
external power supply and do not require power from the 1394 bus.  
VGA  
The VGA connector on the NI 1450 provides connection between the  
NI 1450 and a VGA monitor. Use any standard 15-pin VGA cable to access  
the VGA connector. Figure 3-6 shows the VGA connector location and  
pinout. Table 3-3 lists and describes the VGA connector signals.  
NI 1450 Series Compact Vision System User Manual  
3-8  
ni.com  
 
           
Chapter 3  
LEDs, DIP Switches, and Connectors  
NI 1454  
Compact Vision System  
1
6
11  
5
10 15  
Figure 3-6. VGA Connector  
Table 3-3. VGA Connector Signals  
Pin  
1
Signal Name  
Signal Description  
R
G
Red  
2
Green  
Blue  
3
B
4
NC  
C
No Connect  
5
Common of the NI 1450 main power  
Common of the NI 1450 main power  
Common of the NI 1450 main power  
Common of the NI 1450 main power  
+5V  
6
C
7
C
8
C
9
+5V  
C
10  
11  
12  
13  
14  
15  
Common of the NI 1450 main power  
No Connect  
NC  
SD  
HSync  
VSync  
SC  
Serial Data  
Horizontal Sync  
Vertical Sync  
Serial Clock  
© National Instruments Corporation  
3-9  
NI 1450 Series Compact Vision System User Manual  
 
       
Chapter 3  
LEDs, DIP Switches, and Connectors  
COM1  
COM1 is a high-speed RS-232 (DTE) serial port used for connecting to  
serial devices, such as PLCs, scanners, and lighting devices.  
Note The Serial Port VIs access COM1 as Port 0.  
Figure 3-7 shows the locations of the COM1 DSUB 9-pin connector.  
Refer to Table 3-4 for COM1 signal names and descriptions  
NI 1454  
Compact Vision System  
5 9  
1 6  
Figure 3-7. COM1 DSUB 9-Pin Connector  
Table 3-4. COM1 Connector Signals  
Pin  
1
Signal Name  
DCD  
RXD  
TXD  
DTR  
C
Signal Description  
Data Carrier Detect  
2
Receive Data  
3
Transmit Data  
4
Data Terminal Ready  
Common of the NI 1450 main power  
Data Set Ready  
5
6
DSR  
RTS  
7
Ready to Send  
Clear to Send  
9
RI  
Ring Indicator  
NI 1450 Series Compact Vision System User Manual  
3-10  
ni.com  
 
         
Chapter 3  
LEDs, DIP Switches, and Connectors  
Ethernet  
TRIG 0  
The Ethernet port on the NI 1450 provides connection between the NI 1450  
and the development computer, either directly or through a network port.  
The NI 1450 automatically detects the speed of the connection  
and configures itself accordingly.  
If you are connecting the NI 1450 to the development computer through  
a network port, use a standard Ethernet cable. To connect the NI 1450  
directly to the development computer, use an Ethernet crossover cable.  
The TRIG 0 isolated input on the NI 1450 provides connection to  
external devices, such as proximity sensors and start/stop buttons. For easy  
connection to the TRIG 0 input, use the National Instruments SMB 111  
coaxial cable (part number 763422-01).  
Note Additional isolated inputs are available on the 44-pin DSUB connector.  
Caution These isolated inputs are compatible with 5 V logic if the external circuit meets  
the voltage and current requirements listed in Appendix B, Specifications.  
TRIG 1 and TRIG 2  
You can use the two TTL outputs available on the SMB connectors for  
triggering cameras and external interfaces, such as lighting control units.  
For easy connection to the TTL outputs, use the National Instruments SMB  
111 coaxial cable (part number 763422-01).  
Note Additional TTL outputs are available on the 44-pin DSUB connector.  
Caution Do not connect voltage or current sources to TTL outputs. Doing so could  
damage the NI 1450.  
General-Purpose Digital I/O  
The 44-pin DSUB connector, shown in Figure 3-8, provides access to the  
general-purpose digital inputs and outputs. The general-purpose digital I/O  
available on this connector includes two TTL inputs, eight TTL outputs,  
twelve isolated inputs, and four isolated outputs. For easy connection to the  
digital I/O connector, use the National Instruments digital I/O cable and  
terminal block.  
© National Instruments Corporation  
3-11  
NI 1450 Series Compact Vision System User Manual  
 
                         
Chapter 3  
LEDs, DIP Switches, and Connectors  
For detailed information about digital I/O functionality and recommended  
use cases, refer to Chapter 4, Digital I/O Functionality.  
Note Isolated inputs are compatible with 5 V logic if the external circuit meets the voltage  
and current requirements listed in Appendix B, Specifications.  
NI 1454  
Compact Vision System  
44 30 15  
31 16 1  
Figure 3-8. 44-Pin DSUB Connector  
Table lists pin numbers, signal names, and signal descriptions for the  
44-pin connector on the NI 1450 and the 37-pin terminal block.  
Caution Do not draw more than 500 mA combined from the Viso pins on the 44-pin  
DSUB connector. Do not draw more than 100 mA from each isolated output.  
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals  
44-Pin  
DSUB on  
NI 1450  
37-Pin  
Terminal  
Block  
Pin Number Pin Number  
Signal Name  
TTL Input 0  
C
Primary Function  
Alternate Function  
1
2
1
3
Pulse generator trigger input General-purpose input  
Common of the NI 1450  
main power  
3
4
TTL Output 0  
Watchdog output  
General-purpose output  
NI 1450 Series Compact Vision System User Manual  
3-12  
ni.com  
 
         
Chapter 3  
LEDs, DIP Switches, and Connectors  
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals (Continued)  
44-Pin  
DSUB on  
NI 1450  
37-Pin  
Terminal  
Block  
Pin Number Pin Number  
Signal Name  
TTL Output 1  
C
Primary Function  
Alternate Function  
General-purpose output  
4
5
5
6
Pulse generator output  
Common of the NI 1450  
main power  
6
7
8
7
8
6
TTL Output 2  
TTL Output 3  
C
Pulse generator output  
Pulse generator output  
General-purpose output  
General-purpose output  
Common of the NI 1450  
main power  
9
NC  
17  
18  
19  
35  
34  
9
NC  
No connect  
10  
11  
12  
13  
14  
15  
16  
17  
Viso  
Isolated power  
Ciso  
Isolated common  
General-purpose output  
General-purpose output  
Isolated common  
Input port, Data(0)  
ISO Output 0  
ISO Output 1  
Ciso  
ISO Input 0  
TTL Input 1  
C
2
Pulse generator trigger input General-purpose input  
3
Common of the NI 1450  
main power  
18  
19  
20  
20  
21  
22  
TTL Output 4  
TTL Output 5  
C
Pulse generator output  
General-purpose output  
General-purpose output  
Common of the NI 1450  
main power  
21  
22  
23  
23  
24  
22  
TTL Output 6  
TTL Output 7  
C
General-purpose output  
General-purpose output  
Common of the NI 1450  
main power  
24  
25  
NC  
33  
NC  
No connect  
Viso  
Isolated power  
© National Instruments Corporation  
3-13  
NI 1450 Series Compact Vision System User Manual  
 
Chapter 3  
LEDs, DIP Switches, and Connectors  
Table 3-5. 44-Pin DSUB and 37-Pin Terminal Block Connector Signals (Continued)  
44-Pin  
DSUB on  
NI 1450  
37-Pin  
Terminal  
Block  
Pin Number Pin Number  
Signal Name  
Ciso  
Primary Function  
Isolated common  
Alternate Function  
26  
27  
28  
29  
30  
31  
32  
33  
34  
35  
36  
37  
38  
39  
40  
41  
42  
43  
44  
34  
36  
37  
12  
10  
11  
13  
16  
14  
15  
28  
25  
26  
28  
27  
29  
32  
30  
31  
ISO Output 2  
ISO Output 3  
Ciso  
General-purpose output  
General-purpose output  
Isolated common  
ISO Input 1  
ISO Input 2  
ISO Input 3  
Ciso  
Input port, Data(1)  
Input port, Data(2)  
Input port, Data(3)  
Isolated common  
ISO Input 4  
ISO Input 5  
Ciso  
Input Port, Data(4)  
Input port latch, Data(5)  
Isolated common  
ISO Input 6  
ISO Input 7  
Ciso  
Quadrature encoder Phase A General-purpose input  
Quadrature encoder Phase B General-purpose input  
Isolated common  
ISO Input 8  
ISO Input 9  
Ciso  
Pulse generator trigger input General-purpose input  
General-purpose input  
Isolated common  
General-purpose input  
ISO Input 10  
ISO Input 11  
General-purpose input  
User shutdown  
General-purpose input  
General-purpose input  
NI 1450 Series Compact Vision System User Manual  
3-14  
ni.com  
 
4
Digital I/O Functionality  
This chapter describes the primary functions of the digital inputs and  
outputs on the NI 1450 Series Compact Vision System. This chapter also  
includes guidelines for connecting the digital I/O and for setting up a  
typical NI 1450 system.  
Overview  
The digital I/O functions are accessible through 2 TTL inputs, 10 TTL  
outputs, 13 isolated inputs, and 4 isolated outputs.  
Input signals can be used as triggers, product selection ports, to read  
quadrature encoders, or they can be user-defined. Uses for output signals  
include controlling camera reset and exposure, controlling strobe lighting,  
outputting inspection results, communicating with PLCs, or they can be  
user defined.  
For information about how to use LabVIEW RT to implement specific  
digital I/O functions, refer to the application software documentation and  
examples in the following locations:  
ProgramFiles\NationalInstruments\NI1450Series\Docs  
ProgramFiles\NationalInstruments\LabVIEW7.0\  
examples\NI1450  
The NI 1450 Series Compact Vision System Digital I/O Help is available  
for each example. This help file contains digital I/O reference information  
and instructions for using the LabVIEW FPGA VIs.  
Tip To quickly launch the digital I/O help from an example, press <F1>.  
© National Instruments Corporation  
4-1  
NI 1450 Series Compact Vision System User Manual  
 
           
Chapter 4  
Digital I/O Functionality  
TTL Inputs and Outputs  
TTL is a fast-switching 5 V digital signaling standard commonly used  
for applications that require high precision, such as camera triggering.  
TTL inputs and outputs do not require a separate power supply.  
Caution Do not connect voltage or current sources to TTL outputs. Doing so could  
damage the NI 1450.  
Table 4-1 summarizes the TTL inputs and outputs available on the NI 1450.  
Table 4-1. TTL Inputs and Outputs  
44-Pin DSUB  
on NI 1450  
Pin Number  
37-Pin  
Terminal Block  
Pin Number  
Primary  
Function  
Input or  
Output  
Number  
Available  
Signal Names  
TTL Input 0  
Trigger  
Input  
2
6
1
16  
1
2
TTL Input 1  
Timed Pulse  
Output  
TRIG 1, Pulse 5  
TRIG 2, Pulse 6  
TTL Output 1, Pulse 1  
TTL Output 2, Pulse 2  
TTL Output 3, Pulse 3  
TTL Output 4, Pulse 4  
4
6
7
5
7
8
18  
20  
Watchdog  
Output  
Output  
1
3
TTL Output 0  
3
4
General-Purpose  
TTL Output 5  
TTL Output 6  
TTL Output 7  
19  
21  
22  
21  
23  
24  
Isolated Inputs and Outputs  
The isolated inputs and outputs on the NI 1450 have a separate ground  
reference from the main NI 1450 supply, providing an easy means to  
prevent ground loops that can introduce noise into a system. You can apply  
signals up to 30 V to the isolated inputs. The voltage swing of the isolated  
outputs is determined by the Viso you supply on the connector.  
Table 4-2 summarizes the isolated inputs and outputs available on the  
NI 1450.  
NI 1450 Series Compact Vision System User Manual  
4-2  
ni.com  
 
         
Chapter 4  
Digital I/O Functionality  
Table 4-2. Isolated Inputs and Outputs  
44-Pin DSUB  
on NI 1450  
Pin Number  
37-Pin  
TerminalBlock  
Pin Number  
Input or  
Output  
Number  
Available Signal Names  
Primary Function  
Trigger  
Input  
3
TRIG 0  
ISO Input 5†  
ISO Input 8  
35  
40  
15  
27  
Quadrature Encoder  
Input  
Input  
Input  
1
1
1
ISO Input 6  
ISO Input 7  
37  
38  
25  
26  
External Shutdown  
Control  
ISO Input 11  
44  
31  
Product Selection Port†  
ISO Input 0  
ISO Input 1  
ISO Input 2  
ISO Input 3  
ISO Input 4  
15  
30  
31  
32  
34  
9
10  
11  
13  
14  
General-Purpose  
General-Purpose  
Input  
2
4
ISO Input 9  
ISO Input 10  
41  
43  
29  
30  
Output  
ISO Output 0  
ISO Output 1  
ISO Output 2  
ISO Output 3  
12  
13  
27  
28  
19  
35  
36  
37  
ISO Input 5 can also function as a latch for the product selection port.  
Trigger Inputs  
Trigger inputs are available from both TTL inputs and isolated inputs. You  
can use these trigger inputs to synchronize the NI 1450 with an external  
event, such as the assertion of a signal generated by a proximity sensor or  
a PLC to indicate that an inspection item is passing in front of the camera.  
The NI 1450 uses this input to initiate a timed pulse that can be used for  
camera control, lighting control, encoder pulse counting, and result output  
timing.  
For more information about creating a timed pulse output, refer to the  
Timed Pulse Output section.  
Alternatively, the ISO Input 5 signal can function as a latch input for  
the product selection port, and all five trigger inputs can function as  
general-purpose inputs.  
© National Instruments Corporation  
4-3  
NI 1450 Series Compact Vision System User Manual  
 
       
Chapter 4  
Digital I/O Functionality  
Timed Pulse Output  
The NI 1450 is capable of timed pulse output on six different digital  
outputs, which provides precise control over time-critical signals, such as  
camera exposure. This section describes the various uses for the timed  
pulse output and the parameters you can set to control these outputs.  
Uses for timed pulse output include controlling camera reset and exposure,  
controlling strobe lighting, operating plungers on an assembly line, and  
communicating with PLCs. You can configure the start of the pulse output  
generation to occur from software or from a rising or falling edge of a  
trigger input.  
In addition to controlling the timing of pulse output, you can also configure  
the polarity of the output signal, resulting in a high-true or low-true signal.  
Based on the polarity setting, the output signal asserts after the appropriate  
delay time and de-asserts after the configured pulse width. You can set the  
delay time in microseconds or in quadrature encoder counts from the start  
signal—either a hardware trigger or a software command. Width is always  
configured in microseconds.  
Initiating a Timed Pulse  
Each timed pulse generator has a trigger input that specifies whether to wait  
on a particular trigger input to generate the pulse or to immediately  
generate the pulse when software sets the mode to Start.  
When the pulse generator is configured for a particular trigger input, after  
generating a pulse, it waits for another trigger before generating another  
pulse. When the pulse generator is configured to immediately generate the  
pulse on a software start, after generating a pulse, it immediately generates  
another pulse.  
If the trigger input is set to immediate, the pulse generation occurs as soon  
as the pulse mode is set to start. If the trigger input is set to one of the  
hardware trigger inputs, the timed pulse output waits for an assertion edge  
on the appropriate trigger input. The assertion edge is configurable based  
on the trigger polarity parameter. It then generates one pulse and rearms to  
wait for the next trigger. In either case, the pulse output generation stops  
and resets if the pulse mode parameter is set to stop.  
NI 1450 Series Compact Vision System User Manual  
4-4  
ni.com  
 
       
Chapter 4  
Digital I/O Functionality  
Figure 4-1 shows an output pulse when a trigger is selected.  
Trigger  
Input  
Output  
Pulse  
Figure 4-1. Output Pulse When Trigger Is Selected  
Pulse Modes  
Each pulse generator has a Start and a Stop mode. Configure the pulse  
generator when in Stop mode and then set it to Start mode.  
Pulse Delay  
Pulse delay is the amount of time between a trigger and the first (assertion)  
edge of an output pulse. The pulse delay is configurable in units of  
microseconds or quadrature encoder counts. If configured for  
microseconds, available values are between 10 µs and 4,294,967,295 µs,  
which is 4,294 seconds, or approximately 71 minutes. If the delay is  
configured for quadrature encoder counts, the range of choices is 0 counts  
to 4,294,967,295 counts.  
Pulse Width  
Pulse width is the amount of time between the first (assertion) edge of a  
pulse and the second (deassertion) edge. Pulse width is configurable only  
in microseconds from 10 µs to 4,294,967,295 µs.  
Trigger Polarity  
Each pulse generator can be individually configured for rising or falling  
edge triggering. Even if multiple pulse generators are using the same  
trigger, each can have different polarities.  
Figure 4-2 shows the output of a pulse generator configured to look for a  
rising edge trigger and output a high pulse with a microsecond width and  
delay.  
© National Instruments Corporation  
4-5  
NI 1450 Series Compact Vision System User Manual  
 
                 
Chapter 4  
Digital I/O Functionality  
Delay  
Width  
Trigger  
Pulse  
Figure 4-2. High Pulse with Rising Edge Trigger  
Figure 4-3 shows how to create a high and low pulse train with a  
microsecond delay and width.  
High Pulse Train  
Delay  
Width  
Delay  
Width  
Low Pulse Train  
Software  
Start  
Figure 4-3. High and Low Pulse Trains  
Quadrature Encoder  
The quadrature encoder uses ISO Input 6 for its Phase A input and  
ISO Input 7 for its Phase B input. Encoder speed is limited by the speed of  
the isolated inputs. Each isolated input can change at a maximum rate of  
100 kHz, making the maximum encoder rate 400,000 counts/s.  
The quadrature encoder can also be used as a timebase for the pulse  
generation delay.  
NI 1450 Series Compact Vision System User Manual  
4-6  
ni.com  
 
       
Chapter 4  
Digital I/O Functionality  
Figure 4-4 shows a rising edge trigger and a low pulse with a quadrature  
encoder delay and a microsecond width.  
Trigger  
Low Pulse  
Delay  
Width  
Phase A  
Phase B  
Figure 4-4. Rising Edge Trigger, Low Pulse with a Quadrature Encoder Delay  
Product Selection Port  
The product selection port consists of a group of five isolated digital inputs  
that the software running on the NI 1450 reads simultaneously. You can  
program the NI 1450 to switch between up to 32 (25) inspection sequences  
for different parts on an assembly line.  
Based on the input to the product selection port, you can configure the  
application software to run the appropriate inspection sequence. For  
example, an upstream NI 1450 programmed for part classification might  
drive the product selection port of a downstream NI 1450. Alternatively,  
a PLC with information about which part is being inspected can drive the  
product selection port of the NI 1450.  
Using ISO Input 5 as a Latch  
You can configure the product selection port to use ISO Input 5 as a latch.  
A rising edge on ISO Input 5 can latch the data into a data register on the  
NI 1450. Before each inspection, the software checks the status of the  
product select inputs and reads the most recent value latched into the  
register.  
Note In Vision Builder AI, ISO Input 5 is always designated as a latch.  
© National Instruments Corporation  
4-7  
NI 1450 Series Compact Vision System User Manual  
 
           
Chapter 4  
Digital I/O Functionality  
Table 4-3 lists the product selection ports.  
Table 4-3. Product Selection Port  
Function  
Data(5), rising edge latch  
Data(4)  
External Connection  
ISO Input 5  
ISO Input 4  
ISO Input 3  
ISO Input 2  
ISO Input 1  
ISO Input 0  
Data(3)  
Data(2)  
Data(1)  
Data(0)  
General-Purpose I/O  
General-Purpose Inputs  
General-purpose inputs are available as both TTL and isolated inputs.  
At any time, the software running on the NI 1450 can read these inputs.  
The primary difference between general-purpose inputs and trigger inputs  
is that you cannot use general-purpose inputs to initiate a timed pulse  
generator. In an application, use the general-purpose inputs to get the status  
of the inputs at a given point and not to synchronize the NI 1450 system  
with an external event.  
An example of how to use general-purpose inputs is reading the status of  
a general-purpose input as the first step in your inspection sequence and  
recording that value as part of your inspection.  
General-Purpose Outputs  
General-purpose outputs are also available as both TTL and isolated  
outputs. At any time, the software running on the NI 1450 can drive these  
outputs high or low.  
The primary difference between general-purpose outputs and timed pulse  
outputs is that the timing of general-purpose outputs is controlled by  
software rather than hardware. As a result, timing of general-purpose  
outputs changes as the inspection algorithm changes, which makes  
general-purpose outputs less appropriate than timed outputs for camera  
control, strobe light control, and other applications that require precise  
timing.  
NI 1450 Series Compact Vision System User Manual  
4-8  
ni.com  
 
             
Chapter 4  
Digital I/O Functionality  
An example of using general-purpose outputs is driving a relay that turns  
on an Inspection in Progress light for an operator to see while the inspection  
sequence is running.  
Fault Conditions  
The NI 1450 recognizes the following fault conditions:  
External shutdown, when Shutdown mode is enabled  
Watchdog expiration, when configured for system shutdown  
Overheat  
The behavior of the NI 1450 in the event of a fault condition is dependent  
on configuration settings of the SAFE MODE DIP switch and the  
software-enabled Shutdown mode.  
To resume operation, address the fault condition and reset the NI 1450.  
The following sections describe each fault condition, and Table 4-4  
summarizes how user configuration affects the behavior of the NI 1450  
in the event of a fault condition.  
Shutdown  
Shutdown mode is a software-enabled feature that when activated allows  
an external device to halt the NI 1450 processing operations. Additionally,  
enabling Shutdown mode allows you to specify user-defined shutdown  
states for all fault conditions.  
When Shutdown mode is enabled and the shutdown input signal,  
ISO Input 11, turns off, the NI 1450 registers an external shutdown  
condition. When this fault occurs, the POWER OK LED turns red and all  
system operation halts. To resume operation, you must reset the NI 1450.  
For all fault conditions, outputs operate according to user-defined  
shutdown states. Each TTL output is configurable to drive high, drive low,  
or 3-state, and each isolated output is on/off configurable.  
Because the NI 1450 cannot run VIs in Safe mode, Shutdown mode is  
unavailable when the SAFE MODE DIP switch is turned on.  
Refer to Table 4-4 for a summary of how user configuration affects the  
shutdown condition.  
© National Instruments Corporation  
4-9  
NI 1450 Series Compact Vision System User Manual  
 
     
Chapter 4  
Digital I/O Functionality  
Disabling Shutdown Mode  
To disable Shutdown mode once it is enabled, remove Shutdown mode  
from the software script and reset the NI 1450. To reset the system, press  
the RESET button on the NI 1450 front panel for at least two seconds.  
Note For prototyping when equipment is unavailable, you can wire from Viso to  
ISO Input 11 to simulate external equipment that indicates to the NI 1450 to operate  
normally.  
Watchdog  
The watchdog timer is a software configurable feature that can monitor  
software on the NI 1450 and take action if the software is unresponsive.  
The millisecond counter on the watchdog timer is configurable up to  
65,534 ms, in 1 ms increments, before it expires.  
Configure the watchdog to take one of the following actions when it  
expires:  
Indicator Only—This option sends the expiration signal back to the  
development machine through software. True indicates an expired  
watchdog timer, and False indicates an unexpired watchdog timer. The  
expiration signal that indicates an expired watchdog continues to assert  
until the watchdog is disarmed. Disarming the watchdog resets the I/O.  
Caution Use the Indicator Only option only to test the watchdog. If software becomes  
unresponsive, it cannot be relied upon to send notification to the host.  
TTL Output 0—This option outputs a signal on TTL Output 0. A High  
on TTL Output 0 indicates that the watchdog has expired, and a Low  
indicates that it has not expired. If the watchdog timer has expired, the  
expiration signal will continue to assert until the watchdog is disarmed.  
Note LabVIEW RT supports all watchdog timer options. Vision Builder AI supports only  
the Indicator Only and TTL Output 0 options.  
System Shutdown—This option halts the NI 1450 operation and turns  
the POWER OK LED red. If Shutdown mode is enabled, the outputs  
go to the user-defined shutdown states.  
Table 4-4 summarizes how user configuration affects the shutdown  
condition.  
NI 1450 Series Compact Vision System User Manual  
4-10  
ni.com  
 
     
Chapter 4  
Digital I/O Functionality  
Overheat  
Operating the NI 1450 outside of its temperature specifications may cause  
the NI 1450 to overheat. Refer to Appendix B, Specifications, for  
temperature specifications.  
In the event of an overheat, all NI 1450 operation halts and the  
POWER OK LED turns red. If Shutdown mode is enabled, the outputs  
go to the user-defined shutdown states.  
Table 4-4. Configuration Settings and Effects  
Configuration Settings  
Configuration Effects  
Outputs Change to  
Fault  
Condition  
Safe Mode  
DIP  
Shutdown  
Enabled  
Processing Halts  
User-Defined States  
External  
Shutdown  
On  
Off  
Off  
On  
Off  
Off  
On  
Off  
Off  
N/A  
On  
N/A  
Yes  
No  
No  
Yes  
No  
No  
Yes  
No  
No  
Yes  
No  
Off  
N/A  
On  
Watchdog  
Overheat  
N/A  
Yes  
Yes  
Yes  
Yes  
Yes  
Off  
N/A  
On  
Off  
Considerations When Connecting the Digital I/O  
The isolated trigger inputs on the NI 1450 are current sinking and optically  
isolated. Figure 4-6 shows the trigger input interface circuit.  
Wiring an Isolated Input to a Sourcing Output Device  
You can wire an isolated input to a sourcing output device, as shown in the  
example in Figure 4-5. Refer to Appendix B, Specifications for switching  
thresholds and current requirements.  
Caution Do not apply a voltage greater than 30 VDC to the isolated inputs. Doing so could  
damage NI 1450.  
© National Instruments Corporation  
4-11  
NI 1450 Series Compact Vision System User Manual  
 
                 
Chapter 4  
Digital I/O Functionality  
Sourcing  
Output  
Device  
Viso  
Vcc  
Input  
Current  
Limiter  
Ciso  
NI 1450  
Figure 4-5. Example of Connecting an Isolated Input to a Sourcing Output Device  
Wiring an Isolated Output to an External Load  
The digital output circuit sources current to external loads, as shown in  
the example in Figure 4-6. The maximum output current of this circuit  
is 100 mA.  
Viso  
Vcc  
Digital Output  
Load  
Ciso  
NI 1450  
Figure 4-6. Example of Connecting an Isolated Output to an External Load  
NI 1450 Series Compact Vision System User Manual  
4-12  
ni.com  
 
         
Chapter 4  
Digital I/O Functionality  
Protecting Inductive Loads  
When an inductive load, such as a relay or solenoid, is connected to an  
output, a large counter-electromotive force may occur at switching time  
due to energy stored in the inductive load. This flyback voltage can damage  
the outputs and the power supply.  
To limit flyback voltages at the inductive load, install a flyback diode across  
the load. Mount the flyback diode as close to the load as possible. Use this  
protection method if you connect any of the isolated outputs on the NI 1450  
to an inductive load.  
Figure 4-7 shows an example of using an external flyback diode to protect  
inductive loads.  
Viso  
Vcc  
Digital  
Output  
Load  
Ciso  
External  
Flyback  
Diode for  
NI 1450  
Inductive Loads  
Figure 4-7. Example of Using an External Flyback Diode for Inductive Loads  
Transmission Line Effects  
Transmission line effects can degrade the signals on the I/O cables and  
cause instability. To minimize transmission line effects, use twisted-pair  
wires with a characteristic impedance of 118to connect external signals  
to the 44-pin I/O DSUB connector. Use a 75coaxial cable, such as  
RG-179, to connect to the SMB connectors.  
Figure 4-8 shows connections to the 44-pin DSUB connector and the  
TRIG 0 SMB connector that minimize transmission line effects.  
© National Instruments Corporation  
4-13  
NI 1450 Series Compact Vision System User Manual  
 
             
Chapter 4  
Digital I/O Functionality  
TRIG 1  
SMB  
RG-179  
Coaxial Cable  
75 Ω  
TRIG 1  
Receiving  
Equipment  
44-Pin  
DSUB  
118 Ω  
TTL OUT(0)  
3
2
Receiving  
Equipment  
+5 V  
62 kΩ  
RS  
TTL IN(0)  
16  
17  
Transmitting  
Equipment  
NI 1450  
Figure 4-8. Example Connections  
When connecting to TTL inputs on the NI 1450 device, match the output  
impedance of the transmitting device to the characteristic impedance of the  
cable. For example, if the cable characteristic impedance is 118 , make  
Rs equal to 118 , as shown in Figure 4-8.  
Typical System Setup  
Digital I/O configuration varies depending on the number of cameras your  
system setup requires. You can access the digital I/O through the 44-pin  
DSUB connector or through the 3 SMB connectors.  
The digital I/O capabilities of the SMB connectors on the NI 1450 are ideal  
for typical single-camera applications. The SMB connectors are labeled  
TRIG 0, TRIG 1, and TRIG 2.  
NI 1450 Series Compact Vision System User Manual  
4-14  
ni.com  
 
       
Chapter 4  
Digital I/O Functionality  
Table 4-5 summarizes a typical single-camera configuration.  
Table 4-5. Typical Single-Camera System Setup  
Signal Name  
Signal Type  
Isolated input  
Purpose  
TRIG 0  
Trigger input from  
proximity sensor or  
external device  
TRIG 1  
TRIG 2  
Timed pulse TTL output  
Timed pulse TTL output  
Exposure start and  
control signal to camera  
Strobe light control  
Figure 4-9 shows a typical single-camera setup.  
2
1
NI 1454  
Compact Vision System  
3
4
1
2
Lighting Control Unit  
Lighting Ring  
3
4
Inspection Items  
Proximity Sensor  
Figure 4-9. Typical Single-Camera Setup  
© National Instruments Corporation  
4-15  
NI 1450 Series Compact Vision System User Manual  
 
   
5
Deployment  
This chapter provides guidelines for connecting the NI 1450 Series  
Compact Vision System to a network.  
Connecting Multiple NI 1450s  
An Ethernet connection between the development computer and the  
NI 1450 allows you to assign an IP address, configure the NI 1450,  
download inspection tasks, and remotely monitor an ongoing inspection.  
As with all Ethernet devices, you can connect multiple NI 1450s to the  
same network, as shown in Figure 5-1.  
Figure 5-1. Multiple NI 1450s Connected to the Same Network  
© National Instruments Corporation  
5-1  
NI 1450 Series Compact Vision System User Manual  
 
           
Chapter 5  
Deployment  
To connect multiple NI 1450s to the same network, each NI 1450 must have  
a unique IP address. By default, the configuration utility running on the  
development computer displays each NI 1450 that exists on the subnet.  
To add entries for NI 1450s that exist on other subnets, you must know the  
IP address assigned to each NI 1450. This feature allows remote  
configuration, programming, and monitoring of any NI 1450 not protected  
by a firewall or user password.  
Use the NI 1450 serial number and media access control (MAC) address to  
uniquely identify each unit. The serial number and MAC address are  
printed on the top side of each NI 1450 unit, as shown in Figure 5-2.  
Figure 5-2. Serial Number and MAC Address Label on the NI 1450  
The configuration environment on the host machine displays these values  
in order to distinguish one NI 1450 from another. In addition to  
distinguishing units based on serial number and MAC address, you can also  
assign each NI 1450 a descriptive name in the configuration environment.  
When configured to run an inspection, the NI 1450 can run autonomously  
without connection to the host machine. All image acquisition, inspection,  
decision making, and I/O occurs on the NI 1450 itself. For local monitoring  
of the inspection, you can connect a monitor directly to the Video Out  
connector on the NI 1450, as shown in Figure 5-3.  
NI 1450 Series Compact Vision System User Manual  
5-2  
ni.com  
 
         
Chapter 5  
Deployment  
Figure 5-3. NI 1450 Series Compact Vision System Connected to a Monitor  
At any time, you can reconnect the host machine to the NI 1450 and  
remotely monitor progress.  
© National Instruments Corporation  
5-3  
NI 1450 Series Compact Vision System User Manual  
 
 
A
Troubleshooting  
This appendix provides instructions for troubleshooting the NI 1450.  
Software Configuration Problems  
NI 1450 Does Not Appear in MAX or in Vision Builder AI  
Possible causes and solutions:  
Verify that there is power to the NI 1450 and that both the NI 1450 and  
the development computer are properly connected to the network.  
The POWER OK LED should be lit. When you click Browse,  
the ACT/LINK LED on the NI 1450 should flash to indicate that  
communications are taking place.  
Certain network devices, such as switches and routers, may filter out  
the directed UDP broadcast packets the development computer uses to  
communicate with unconfigured NI 1450s, even if these devices are on  
the same subnet. You can test this by connecting the development  
computer and the NI 1450 directly to a network port or by using a  
crossover cable between the NI 1450 and the development computer.  
If the NI 1450 appears in the software using this configuration, but not  
when connected to the larger network, the network is not properly  
communicating with the NI 1450. Consult your network administrator.  
The NI 1450 may have been configured on another network and then  
moved to the current network. Reconfigure the NI 1450 on the current  
network.  
The grade of cable you are using may be insufficient for the speed of  
your network, or you may be using the wrong type of Ethernet cable.  
Use a straight-through cable when connecting to network hardware.  
Use a crossover cable to connect directly to the development computer.  
© National Instruments Corporation  
A-1  
NI 1450 Series Compact Vision System User Manual  
 
       
Appendix A  
Troubleshooting  
Another device on the network is using the IP address assigned to the  
NI 1450. Either remove or reconfigure the other device, or reconfigure  
the NI 1450 IP address.  
The DIP switch settings on the NI 1450 may be invalid, such as all  
switches set to the ON position. Change the DIP switch settings and  
reset the NI 1450 by pressing the RESET button on the NI 1450 for at  
least two seconds.  
Error When Configuring IP Settings for the NI 1450  
The NI 1450 cannot be configured if the subnet mask or IP address is  
incompatible with the network settings. Verify that the values entered for  
Status of NI 1450 in MAX and Vision Builder AI is NoSoftware  
Installed  
Install application and driver software on the NI 1450. Refer to Chapter 2,  
Setup and Configuration, for installation instructions.  
No Camera Found  
Verify that the total power requirements of the cameras do not exceed the  
power the NI 1450 provides on the 1394 bus. Verify that the cameras  
comply with the IIDC Digital 1394-based Camera Specification,  
Version 1.30.  
Hardware Errors  
POWER OK LED is Red  
The NI 1450 is in one of the following fault conditions:  
Shutdown enabled in software, but ISO Input 11 is not turned on  
ISO Input 11 is used to indicate to the NI 1450 that all external  
devices are ready for operation. When ISO Input 11 turns off, the  
NI1450 shuts down. Either turn on ISO Input 11 with an external  
source and reset the NI 1450, or disable Shutdown in the software.  
Note For prototyping when external equipment is unavailable, Viso can be wired to ISO  
NI 1450 Series Compact Vision System User Manual  
A-2  
ni.com  
 
   
Appendix A  
Troubleshooting  
Watchdog expired while configured for the system shutdown option  
Determine if the interval between watchdog resets is insufficient  
for the processing time. Check the software to determine if it  
could monopolize the processor.  
Processor overheat  
Make sure you are operating the NI 1450 in compliance with the  
temperature specifications in Appendix B.  
Refer to Table 4-4, Rising Edge Trigger, Low Pulse with a Quadrature  
Encoder Delay, for more information about fault conditions.  
Cannot Drive Isolated Outputs  
Verify that you have power connected to Viso and Ciso inputs on the  
NI 1450 power connector and that the contact at the screw terminals is  
solid. Because these outputs are electrically isolated from the NI 1450  
main supply, they require power in addition to the NI 1450 main supply.  
Runaway Startup Application  
If a runaway startup application causes the NI 1450 to become  
unresponsive, power off the NI 1450 and then reboot it with either the NO  
APP switch or the SAFE MODE switch in the ON position. Rebooting the  
NI 1450 with the NO APP switch enabled prevents any VIs from running  
at startup. Rebooting with the SAFE MODE switch enabled starts the  
NI 1450 in Safe mode and does not launch the embedded LabVIEW RT  
engine. Refer to the DIP Switches section of Chapter 3, LEDs, DIP  
Switches, and Connectors, for more information about these DIP switches.  
System Displays a MissingOperatingSystemError at Startup  
The NI 1450 was powered on or reset with all DIP switches in the ON  
position. Set the DIP switches to the off position, and reset the NI 1450  
by pressing the RESET button for at least two seconds.  
LED Error Indications  
The NI 1450 indicates specific error conditions by flashing the orange  
STATUS LED a specific number of times. Table A-1 describes the  
STATUS LED flashing sequences and the corresponding error condition.  
© National Instruments Corporation  
A-3  
NI 1450 Series Compact Vision System User Manual  
 
 
Appendix A  
Troubleshooting  
Table A-1. STATUS LED Flashes and Corresponding Error Conditions  
Number  
of Flashes  
Error Condition  
0 (stays lit)  
The NI 1450 has detected an internal error. Contact  
National Instruments for assistance.  
1
2
The NI 1450 is unconfigured.  
The NI 1450 has detected an error in its software. This  
usually occurs when an attempt to upgrade the firmware  
is interrupted. Repeat the firmware upgrade process.  
3
The NI 1450 is in Safe mode.  
4 (or more) The NI 1450 has detected an unrecoverable error.  
Contact National Instruments for assistance.  
POWER OK LED Is Not Lit When the NI 1450 is  
Powered on  
If the power supply is properly connected to the NI 1450, but the POWER  
OK LED does not light up, check that the power supply is 24 VDC 10%  
and within the specifications outlined in Appendix B, Specifications. Using  
a power supply that is not within these specifications might result in an  
unresponsive or unstable system and could damage the NI 1450.  
NI 1450 Series Compact Vision System User Manual  
A-4  
ni.com  
 
 
B
Specifications  
This appendix lists the specifications of the NI 1450 Series Compact Vision  
System. These specifications are typical at 25 °C, unless otherwise noted.  
Power Requirements  
Main supply voltage............................... 24 VDC 10%  
Power (excluding cameras)............. 12 W, typical  
22 W, maximum  
1394 bus power............................... 18 W, maximum  
(shared by all three ports)  
Isolated supply1 ...................................... 5 to 30 VDC  
Memory  
Network  
SDRAM ................................................. 128 MB  
Nonvolatile storage ................................ 32 MB  
Network interface................................... 10BaseT and 100BaseTX  
Ethernet  
Compatibility ......................................... IEEE802.3  
Communication rates ............................. 10 Mbps, 100 Mbps,  
auto-negotiated  
Maximum cabling distance .................... 100 m/segment  
1
Do not draw more than 500 mA combined from the Viso pins on the 44-pin DSUB connector. Do not draw more than 100 mA  
from each isolated output.  
© National Instruments Corporation  
B-1  
NI 1450 Series Compact Vision System User Manual  
 
               
Appendix B  
Specifications  
TTL Inputs and Outputs  
Digital logic levels  
Level  
Minimum  
0 V  
Maximum  
0.5 V  
5 V  
Input low voltage (VIL)  
Input high voltage (VIH)  
2.2 V  
2.4 V  
Output low voltage (VOL), at 5 mA  
Output high voltage (VOH), at 5 mA  
0.4 V  
TTL Inputs  
Number of channels................................2  
Maximum pulse rate ...............................2 MHz  
Minimum pulse detected ........................200 ns  
TTL Outputs  
Number of channels................................10  
Output voltage range ..............................0 V to 5 V  
Maximum pulse rate ...............................2 MHz  
Optically Isolated Inputs and Outputs  
Isolated (Current Sinking) Inputs  
Number of channels................................13  
Input voltage range .................................0 V to 30 V  
Input ON voltage.............................3.5 V to 30 V  
Input OFF voltage............................0 V to 2 V  
Turn-on current.......................................4 mA, typical;  
7.1 mA, minimum;  
14 mA, maximum  
(current limited)  
NI 1450 Series Compact Vision System User Manual  
B-2  
ni.com  
 
     
Appendix B  
Specifications  
Maximum pulse rate............................... 100 kHz  
Minimum pulse detected........................ 10 µs  
Reverse polarity protection .................... yes  
Isolated (Current Sourcing) Outputs  
Number of channels ............................... 4  
On-state voltage range ........................... 5 V to 30 V, maximum  
Maximum on-state voltage  
drop from Viso................................ 1.2 V at 100 mA  
Output current ........................................ 100 mA, maximum  
Maximum pulse rate............................... 10 kHz (maximum load resistance  
100 k)  
Minimum pulse generated...................... 100 µs  
Reverse polarity protection .................... Yes  
IEEE 1394  
Number of ports ..................................... 3  
Speed...................................................... 100, 200, or 400 Mbps  
Available bus power............................... Refer to the Power Requirements  
section  
Physical Characteristics  
Dimensions............................................. 10.2 × 12.7 × 6.4 cm  
(4 × 5 × 2.5 in.)  
Weight.................................................... 977 g (2 lbs 2.4 oz)  
© National Instruments Corporation  
B-3  
NI 1450 Series Compact Vision System User Manual  
 
     
Appendix B  
Specifications  
Environmental  
The NI 1450 is intended for indoor use only.  
Operating temperature,  
vertical mounting position......................0 to 55 °C  
Operating temperature,  
all other positions....................................0 to 45 °C  
Storage temperature................................–55 to 85 °C  
Humidity.................................................10 to 90% RH, noncondensing  
Pollution Degree.....................................2  
Safety  
The NI 1450 Series Compact Vision System meets the requirements of the  
following standards for safety and electrical equipment for measurement,  
control, and laboratory use:  
EN 61010-1, IEC 61010-1  
UL 3121-1, UL 61010C-1  
CAN/CSA C22.2 No. 1010.1  
Note For UL and other safety certifications, refer to the product label or visit ni.com.  
Electromagnetic Compatibility  
Emissions................................................EN 55011 Class A at 10 m  
FCC Part 15A above 1 GHz  
Immunity ................................................EN 61326:1997 + A2:2001,  
Table 1  
EMC/EMI ...............................................CE, C-Tick, and FCC Part 15  
(Class A) Compliant  
Note For full EMC compliance, you must operate this device with shielded cabling.  
NI 1450 Series Compact Vision System User Manual  
B-4  
ni.com  
 
     
Appendix B  
Specifications  
CE Compliance  
This product meets the essential requirements of applicable European  
Directives, as amended for CE marking, as follows:  
Low-Voltage Directive (safety) ............. 73/23/EEC  
Electromagnetic Compatibility  
Directive (EMC) .................................... 89/336/EEC  
Note Refer to the Declaration of Conformity (DoC) for this product for any additional  
regulatory compliance information. To obtain the DoC for this product, click Declarations  
of Conformity Information at ni.com/hardref.nsf/. This Web site lists the DoCs by  
product family. Select the appropriate product family, followed by your product, and a link  
appears to the DoC in Adobe Acrobat format. Click the Acrobat icon to download or read  
the DoC.  
© National Instruments Corporation  
B-5  
NI 1450 Series Compact Vision System User Manual  
 
C
Technical Support and  
Professional Services  
Visit the following sections of the National Instruments Web site at  
ni.comfor technical support and professional services:  
Support—Online technical support resources include the following:  
Self-Help Resources—For immediate answers and solutions,  
visit our extensive library of technical support resources available  
in English, Japanese, and Spanish at ni.com/support. These  
resources are available for most products at no cost to registered  
users and include software drivers and updates, a KnowledgeBase,  
product manuals, step-by-step troubleshooting wizards,  
conformity documentation, example code, tutorials and  
application notes, instrument drivers, discussion forums,  
a measurement glossary, and so on.  
Assisted Support Options—Contact NI engineers and other  
measurement and automation professionals by visiting  
ni.com/ask. Our online system helps you define your question  
and connects you to the experts by phone, discussion forum,  
or email.  
Training—Visit ni.com/trainingfor self-paced tutorials, videos,  
and interactive CDs. You also can register for instructor-led, hands-on  
courses at locations around the world.  
System Integration—If you have time constraints, limited in-house  
technical resources, or other project challenges, NI Alliance Program  
members can help. To learn more, call your local NI office or visit  
ni.com/alliance.  
Declaration of Conformity (DoC)—A DoC is our claim of  
compliance with the Council of the European Communities using the  
manufacturer’s declaration of conformity. This system affords the user  
protection for electronic compatibility (EMC) and product safety. You  
can obtain the DoC for your product by visiting  
ni.com/hardref.nsf.  
© National Instruments Corporation  
C-1  
NI 1450 Series Compact Vision System User Manual  
 
                         
Appendix C  
Technical Support and Professional Services  
Calibration Certificate—If your product supports calibration, you  
can obtain the calibration certificate for your product at  
ni.com/calibration.  
If you searched ni.comand could not find the answers you need, contact  
your local office or NI corporate headquarters. Phone numbers for our  
worldwide offices are listed at the front of this manual. You also can visit  
the Worldwide Offices section of ni.com/niglobalto access the branch  
office Web sites, which provide up-to-date contact information, support  
phone numbers, email addresses, and current events.  
NI 1450 Series Compact Vision System User Manual  
C-2  
ni.com  
 
   
Glossary  
Symbol  
Prefix  
micro  
milli  
Value  
106  
103  
103  
µ
m
k
kilo  
M
mega  
106  
Symbols  
°C  
Degrees Celsius.  
A
AC  
Alternating current.  
acquisition window  
The image size specific to a video standard or camera resolution.  
B
b
Bits.  
B
Bytes.  
buffer  
Temporary storage for acquired data.  
D
DAQ  
Data acquisition. (1) Collecting and measuring electrical signals from  
sensors, transducers, and test probes or fixtures and inputting them to a  
computer for processing; (2) Collecting and measuring the same kinds of  
electrical signals with A/D or DIO boards plugged into a computer, and  
possibly generating control signals with D/A and/or DIO boards in the  
same computer.  
DC  
Direct current.  
DCAM-compliant  
Cameras that comply with the IIDC 1394-based Digital Camera  
Specification, Version 1.30.  
© National Instruments Corporation  
G-1  
NI 1450 Series Compact Vision System User Manual  
 
   
Glossary  
DIP switch  
DNS  
Dual Inline Package switch.  
Domain Name System.  
E
Ethernet crossover cable Category 5 or Category 6 cable used for direct connection between the  
development computer and the Real-Time target.  
Ethernet cable, standard Category 5 or Category 6 Ethernet cable used to connect a Real-Time target  
to a network port.  
F
FTP  
File Transfer Protocol.  
H
Hz  
Hertz. Frequency in units of one cycle per second.  
Institute of Electrical and Electronics Engineers. A standard-setting body.  
I
IEEE  
interrupt  
A computer signal indicating that the CPU should suspend its current task  
to service a designated activity.  
interrupt level  
The relative priority at which a device can interrupt.  
Input/Output.  
I/O  
IRQ  
Interrupt request. See interrupt.  
L
LED  
Light-emitting diode.  
M
m
Meters.  
MAC  
Media access control. The MAC address uniquely identifies each unit  
connected to a network.  
NI 1450 Series Compact Vision System User Manual  
G-2  
ni.com  
 
Glossary  
N
NI-IMAQ  
Driver software for National Instruments image acquisition (IMAQ)  
hardware.  
P
PG  
Pulse Generation.  
PLC  
Programmable Logic Controller. An industrial computer used for factory  
automation, process control, and manufacturing systems.  
POST  
Power-On Self Test.  
R
ROI  
Region of interest. A hardware-programmable rectangular portion of the  
acquisition window.  
RS-232  
Standard electrical interface for serial data communications.  
S
subnet  
A set of systems whose IP addresses are configured such that they can  
communicate directly with one another. Data will not flow through an  
intermediate router.  
T
TCP  
Transmission Control Protocol. A set of standard protocols for  
communicating across a single network or interconnected set of networks.  
TCP is for high-reliability transmissions.  
U
UDP  
User Datagram Protocol. A set of standard protocols for communicating  
across a single network or interconnected set of networks. UDP is for  
low-overhead transmissions.  
© National Instruments Corporation  
G-3  
NI 1450 Series Compact Vision System User Manual  
 
Glossary  
V
V
Volts.  
VDC  
VI  
Volts direct current.  
Virtual Instrument. A combination of hardware and/or software elements,  
typically used with a PC, that has the functionality of a classic stand-alone  
instrument.  
W
W
Watts.  
NI 1450 Series Compact Vision System User Manual  
G-4  
ni.com  
 
Index  
IEEE 1394, 3-8  
SMB, 3-6  
TRIG 0, 3-11  
Numerics  
100 Mbps LED, 3-2  
TRIG 1, 3-11  
TRIG 2, 3-11  
VGA, 3-8  
A
accessories, 2-2  
contacting National Instruments, C-2  
conventions used in the manual, v  
customer  
application software  
installing LabVIEW, 2-15  
installing Vision Development  
Module, 2-15  
technical support, C-1  
B
DCAM specification, 1-1, G-1  
Declaration of Conformity, C-1  
deployment, 5-1  
C
cables  
development computer  
crossover, 2-3  
digital I/O, 2-2  
Ethernet crossover, 2-2  
Ethernet standard, 2-2  
SMB to BNC, 2-3  
installing application software, 2-13  
installing LabVIEW, 2-15  
Module, 2-15  
characteristic impedance, 4-14  
calibration certificate, C-2  
camera  
available bandwidth, 1-3  
connecting to the NI 1450, 2-8  
digital camera specification, 1-1  
video formats, 1-3  
COM1, connector signals, 3-10  
connecting multiple systems  
subnet, 5-2  
diagnostic resources, C-1  
digital camera specification, 1-1  
digital I/O  
44-pin DSUB, 3-11  
cable, D44, 3-11  
connection considerations, 4-11  
connector (table), 3-6  
connector pinout (diagram), 3-12  
connector signals (table), 3-12  
external load, wiring, 4-12  
general-purpose, 3-11  
connectors, 3-6  
COM1, 3-10  
Ethernet, 3-11  
© National Instruments Corporation  
I-1  
NI 1450 Series Compact Vision System User Manual  
 
 
Index  
isolated I/O, 4-2  
signal functionality, 4-1  
(diagram), 4-12  
flyback voltage, flyback diode, 4-13  
frequently asked questions, C-1  
transmission line effects, 4-13  
TTL, 4-2  
DIP switches  
IP RESET, 3-5  
location, 3-4  
accessories, 2-2  
NO APP, 3-5  
overview, 3-3  
SAFE MODE, 3-4  
connecting a camera, 2-8  
computer, 2-11  
documentation  
hardware, 2-3  
LabVIEW RT, 2-3  
online library, C-1  
overview, 2-6  
Vision Builder AI, 2-3  
Vision Development Module, 2-3  
required items, 2-1  
subnets, 2-8  
drivers  
help  
instrument, C-1  
software, C-1  
technical support, C-1  
E
I
specifications, B-4  
environmental specifications, B-4  
error indications (table), A-3  
Ethernet  
IEEE 1394  
cable, 2-1  
connector, 3-8  
licensing, 1-1  
IEEE 1394 bus, available power, B-1  
IEEE 1394 camera  
connecting to NI 1450, 2-8  
digital camera specification, 1-1  
IEEE 1394a (table), 3-6  
image acquisition  
connecting NI 1450  
diagram, 2-12  
connector, 3-11  
crossover cable, 2-3  
LINK LED, 3-2  
port (table), 3-6  
example code, C-1  
external load, wiring, 4-12  
LabVIEW RT, 2-17  
Vision Builder AI, 2-14  
inductive loads, protecting, 4-13  
NI 1450 Series Compact Vision System User Manual  
I-2  
ni.com  
 
Index  
inputs  
general-purpose, 4-8  
National Instruments  
calibration certificate, C-2  
customer education, C-1  
Declaration of Conformity, C-1  
professional services, C-1  
system integration services, C-1  
technical support, C-1  
worldwide offices, C-2  
network  
TTL, 4-2  
instrument drivers, C-1  
IP address  
configuring with Vision Builder AI, 2-13  
configuring with Vision Development  
Module, 2-16  
isolated input, sourcing output device, 4-11  
isolated output, external load, 4-11  
connecting, 2-7  
subnet, 2-8  
K
KnowledgeBase, C-1  
computer, 2-11  
L
connectors (table), 3-6  
DIP switches, overview, 3-3  
IEEE 1394 connector, 3-8  
optional hardware, 2-2  
overview, 1-1  
label, MAC address, serial number, 5-2  
LabVIEW RT  
acquiring images, 2-17  
description, 1-5  
installing, 2-15  
parts locator, 1-2  
LabVIEW, installing, 2-15  
latch, 4-7  
LEDs  
NI-IMAQ for IEEE 1394 Cameras  
installing, 2-13  
diagram, 3-1  
STATUS, 3-2  
online technical support, C-1  
outputs  
M
MAC address, 5-2  
TTL, 4-2  
MAX, 2-16  
Measurement & Automation Explorer  
(MAX), 2-16  
media access control address, 5-2  
memory, specifications, B-1  
mounting kits, part numbers, 2-2  
P
parts locator, 1-2  
phone technical support, C-2  
physical characteristic specifications, B-3  
© National Instruments Corporation  
I-3  
NI 1450 Series Compact Vision System User Manual  
 
Index  
POWER OK LED, 3-2  
power requirements specifications, B-1  
power supply  
software  
application, 1-4, 1-5  
installing on development computer, 2-13  
NI-IMAQ for IEEE 1394 Cameras, 1-4  
software choices, 1-4  
software drivers, C-1  
sourcing output device  
wiring, 4-11  
connector, 3-6, 3-7  
connector (table), 3-6  
desktop, 2-1  
earth ground, 3-7  
separate main supply, 2-10  
terminals (table), 3-7  
wiring power, 2-9  
diagram, 4-12  
specifications  
wiring power (diagram), 2-10  
product selection port, 4-7  
(table), 4-8  
electromagnetic compatibility, B-4  
environmental, B-4  
memory, B-1  
professional services, C-1  
programming examples, C-1  
pulse delay, 4-5  
network, B-1  
optically isolated inputs, B-2  
optically isolated outputs, B-3  
outputs, B-3  
pulse modes, 4-5  
pulse width, 4-5  
physical characteristics, B-3  
power requirements, B-1  
safety, B-4  
TTL inputs, B-2  
TTL outputs, B-2  
Q
quadrature encoder, 4-6  
STATUS LED, 3-2  
STATUS LED error indications (table), A-3  
R
required hardware, 2-1  
S
SAFE MODE DIP switch, 3-4  
safety information, 2-4  
safety specifications, B-4  
serial number, 5-2  
technical support, C-1  
telephone technical support, C-2  
termincal block, 37-pin (table), 3-12  
timed pulse output, 4-4  
timed pulse output, initiating, 4-4  
training, customer, C-1  
transmission line effects, 4-13  
TRIG 0  
serial port, connector (table), 3-6  
setup  
multiple NI 1450s, 5-1  
typical single-camera, 4-14  
shutdown, 4-9  
cable, 3-11  
connector (table), 3-6  
shutdown, disabling, 4-10  
NI 1450 Series Compact Vision System User Manual  
I-4  
ni.com  
 
TRIG 1  
cable, 3-11  
connector (table), 3-6  
TRIG 2  
Vision Development Module  
configuring IP address, 2-16  
description, 1-5  
installing, 2-15  
cable, 3-11  
connector (table), 3-6  
trigger input lines, 4-3  
trigger polarity, 4-5  
troubleshooting  
Web  
professional services, C-1  
technical support, C-1  
worldwide technical support, C-2  
hardware, A-2  
software, A-1  
STATUS LED error indications  
(table), A-3  
V
VGA  
connector (table), 3-6  
connector signals (table), 3-9  
VGA connector (diagram), 3-9  
video formats, 1-3  
Vision Builder AI  
configuring IP address, 2-13  
description, 1-5  
image acquisition, 2-14  
installing, 2-13  
© National Instruments Corporation  
I-5  
NI 1450 Series Compact Vision System User Manual  
 

Miele Clothes Dryer 04 950 470 User Manual
Miele Clothes Dryer T8866WP User Manual
Miele Washer 05 960 780 User Manual
Multiquip Portable Generator GA 25H User Manual
Multiquip Projection Television GB114BW User Manual
Nady Systems Music Mixer DSM 1 User Manual
NEC Drill 2601 20 User Manual
Onkyo DVD Player SP404 User Manual
Optiview Security Camera WIPCAM User Manual
Oster Toaster 118341 User Manual